Teledyne 4060 manual Range Function, Analog Output Adjustment

Page 55

Series 4060

Operation

 

 

 

4.6.10The Range Function

________________________________________________

-> Range Select: Man

Man Rng: R1

R1: 0-100.00ppb

R2: 0- 1.00ppm

R3: 0- 10.00ppm

________________________________________________

The value of the upper limit of the currently selected range (AUTO or MANUAL) affects the voltage and current of the analog outputs on the rear of the analyzer as well as contact closures indicating which range is current. The analog outputs from the 50 –pin Equipment connector (see Section 3.3.2.4 and Table 3-1) represent the proportion of the analysis concentration to the currently selected range limit.

MANUAL/AUTO RANGE SELECT: In Manual Range Select, the range indicated on the second line 'Man Rng:' is maintained as the current range of operation. In Auto Range select, the software determines the smallest range whose limit is larger than the analysis concentration. Automatic Range selection in Auto Range is governed by hysteresis such that the concentration must fall below 90% of the limit of the next smaller range before the current range will be set to the lower range.

RANGE LIMITS: Ranges may be set as low as 100.0 ppb and as high as 100.0 ppm. Range 3 can only be set at a value greater than Range 2. Range 2 can only be set at a value greater than Range 1.

4.6.11 Analog Output Adjustment

The software in this instrument provides a way to enter to manage the analog output of the analyzer. To access the offset function:

<UP/DOWN >From the MAIN MENU, scroll to the ANALOG ADJUST function using the UP/DOWN keys.

<Enter> Pressing <Enter> activates the function and takes you to the next screen.

Teledyne Analytical Instruments

55

Image 55
Contents Series Teledyne Analytical Instruments SeriesTeledyne Analytical Instruments Iii Safety MessagesSeries Combustible GAS Usage Additional Safety InformationSeries Teledyne Analytical Instruments Vii Procedure for Removal of Internal Inaccessible Shock HazardsViii Table of Contents Operation Maintenance & Troubleshooting Appendix B AppendixTeledyne Analytical Instruments Xiii List of FiguresList of Tables Introduction Main Features of the AnalyzerSeries Introduction Analyzer Description Principle of OperationIntroduction Series Model 4060 Front Panel Interface Applications Series Operational Theory Operational TheoryIntroduction Modes of OperationOperational Theory Series Analyzer SubsystemsSample System Gas Flow Control System Internal Temperature ControllersFlame Ionization Detection Cell Fuel and Blanket Air SystemsTypical Piping Diagram for the Model Detection Cell Flame Ionization Cell Electrometer-AmplifierFlame Guard Circuit Anode Power SupplyFlame Ignition Circuit Mounting the Analyzer InstallationSeries Installation Unpacking the AnalyzerElectronic Connections Installation SeriesElectrical Power Connections User ConnectionsModel 4060 Rear Panel with Optional Gas Manifold Equipment Interface Connector Pin Arrangement Pin Function Threshold Alarm System Alarm Remote Calibration Connections Range ID Relay Connections 2.10 RS-232 Port Pin out of 50 pin D-Sub ConnectorGas Connections Commands via RS-232 InputGas Connections Less than 0.5 PPM Placing the System in Operation Equipment OperationSeries Operation Activating the Support Gases Preliminary Power-Off Check ListOperation Series Span Gas Flame Ignition1 Air Carrier GasIgnition and/or Flame Guard Circuit Failure Sample PumpAnalyzer Operation Verification of the Flame Guard CircuitSeries Operation Style Conventions Default ParametersEnter Navigation and Data EntryMenu Structure EscapeMain Menus Span ALT-SPAN System Self-Diagnostic TestModel Screen Alarms Function AL-1 Defeated AL-1 High AL-1 NON-FAILSAFE AL-1 NON-LATCHING Operation Series Range Function Analog Output AdjustmentLinearization Setting up an AUTO-CALChanging Stream Autocal Password Password Protection Group SetupTiming Entering a Password Enter NEW Password Screen Logging Out Repeat Password Entry ScreenStandby Series Operation Operation Series Teledyne Analytical Instruments Series Maintenance Maintenance & TroubleshootingMeasuring Circuit Electrical Checks Maintenance SeriesAnode Voltage Check Electronic Stability Printed Circuit Board ReplacementCollector Cable Temperature Control Electronic CheckIgnition and/or Flame Guard Circuit Checks Sampling System Printed Circuit Board Descriptions Flame Guard and Anode Power Supply PCBProportional Temperature Controller PCB Electrometer-Amplifier PCB Maintenance Series Appendix Specifications and Initial SettingsSeries Appendix Appendix Series Recommended Spare Parts ListAnalytical Instruments PC Board Assemblies Drawing ListAppendix Series Teledyne Analytical Instruments Appendix B B1 Addendum and Testing Results