Teledyne 4060 manual Procedure for Removal of Internal Inaccessible Shock Hazards

Page 7

Series 4060

3.External hazardous voltages which may be connected to the Analyzer alarm relay connections.

Procedure for Removal of Internal Inaccessible Shock Hazards

CAUTION: SERVICING OR MAINTENANCE OF THE ANALYZER SHOULD ONLY BE DONE BY SUITABLE TRAINED PERSONNEL. TO AVOID THESE INACCESSIBLE HAZARDOUS VOLTAGES WHEN SERVICING THE SERIES 4060, PERFORM EACH OF THE FOLLOWING STEPS, IN THE ORDER GIVEN, BEFORE SERVICING BEGINS:

1.Switch off the power to the Analyzer and remove the main (line) power cord from the Analyzer.

2.Remove all external voltages from the connections to the alarm contacts.

3.Wait one minute.

4.Discharge the anode supply voltage.

a.Connect one end of an insulated (to 1000 VDC or more) clip lead to Analyzer chassis ground (the standoff for the upper right corner of the mother PCB).

b.Put one end of a 500V rated 1000 ohm resistor in the other end of the clip lead.

c.Check the voltage between chassis ground (the standoff for the upper right corner of the mother PCB) and the top side of R2 at PCB number B74671. It should be between -5VDC and +5VDC. If is in that range, the inaccessible hazardous voltage removal procedure is completed, if not repeat steps 4.a and 4.b.

If it is absolutely necessary to work inside the instrument with power on, use the ONE HAND RULE:

Work with one hand only.

Keep the other hand free without contacting any other object. This reduces the possibility of a ground path through the body in case of accidental contact with hazardous voltages.

Teledyne Analytical Instruments

vii

Image 7
Contents Series Teledyne Analytical Instruments SeriesTeledyne Analytical Instruments Iii Safety MessagesSeries Combustible GAS Usage Additional Safety InformationSeries Teledyne Analytical Instruments Vii Procedure for Removal of Internal Inaccessible Shock HazardsViii Table of Contents Operation Maintenance & Troubleshooting Appendix B AppendixTeledyne Analytical Instruments Xiii List of FiguresList of Tables Introduction Main Features of the AnalyzerSeries Introduction Analyzer Description Principle of OperationIntroduction Series Model 4060 Front Panel Interface Applications Series Operational Theory Operational TheoryIntroduction Modes of OperationOperational Theory Series Analyzer SubsystemsSample System Gas Flow Control System Internal Temperature ControllersFlame Ionization Detection Cell Fuel and Blanket Air SystemsTypical Piping Diagram for the Model Detection Cell Flame Ionization Cell Electrometer-AmplifierFlame Guard Circuit Anode Power SupplyFlame Ignition Circuit Mounting the Analyzer InstallationSeries Installation Unpacking the AnalyzerElectronic Connections Installation SeriesElectrical Power Connections User ConnectionsModel 4060 Rear Panel with Optional Gas Manifold Equipment Interface Connector Pin Arrangement Pin Function Threshold Alarm System Alarm Remote Calibration Connections Range ID Relay Connections 2.10 RS-232 Port Pin out of 50 pin D-Sub ConnectorGas Connections Commands via RS-232 InputGas Connections Less than 0.5 PPM Placing the System in Operation Equipment OperationSeries Operation Activating the Support Gases Preliminary Power-Off Check ListOperation Series Span Gas Flame Ignition1 Air Carrier GasIgnition and/or Flame Guard Circuit Failure Sample PumpAnalyzer Operation Verification of the Flame Guard CircuitSeries Operation Style Conventions Default ParametersEnter Navigation and Data EntryMenu Structure EscapeMain Menus Span ALT-SPAN System Self-Diagnostic TestModel Screen Alarms Function AL-1 Defeated AL-1 High AL-1 NON-FAILSAFE AL-1 NON-LATCHING Operation Series Range Function Analog Output AdjustmentLinearization Setting up an AUTO-CALChanging Stream Autocal Password Password Protection Group SetupTiming Entering a Password Enter NEW Password Screen Logging Out Repeat Password Entry ScreenStandby Series Operation Operation Series Teledyne Analytical Instruments Series Maintenance Maintenance & TroubleshootingMeasuring Circuit Electrical Checks Maintenance SeriesAnode Voltage Check Electronic Stability Printed Circuit Board ReplacementCollector Cable Temperature Control Electronic CheckIgnition and/or Flame Guard Circuit Checks Sampling System Printed Circuit Board Descriptions Flame Guard and Anode Power Supply PCBProportional Temperature Controller PCB Electrometer-Amplifier PCB Maintenance Series Appendix Specifications and Initial SettingsSeries Appendix Appendix Series Recommended Spare Parts ListAnalytical Instruments PC Board Assemblies Drawing ListAppendix Series Teledyne Analytical Instruments Appendix B B1 Addendum and Testing Results