Emerson oxymitterdr hazardous area in-siu oxygen probe, ib-106-340cdr instruction manual G02 G04

Page 33

Hazardous Area Oxymitter DR

Instruction Manual

IB-106-340CDR Original Issue January, 2002

c.WESTINGHOUSE MODEL 218A ELEC- TRONICS SETUP

Before beginning operation, it is important that the probe heater setpoint of the existing electronics be changed to support the Haz- ardous Area Oxymitter DR probe. To con- vert the Model 218A Digital Electronics Package for use with the DR probe, an EPROM change is necessary. Remove Main PCB and check back of board to iden- tify unit as G02 or G04. The replacement EPROM needed is as identified below:

 

G02

G04

United States

1M03192G01

1M02982G01

United Kingdom

1M03192G02

1M02982G02

Germany

1M03192G03

1M02982G03

France

1M03192G04

1M02982G04

Italy

1M03192G05

1M02982G05

NOTE

The replacement EPROM when using a multiprobe averager unit is

1M02982G10.2

To replace the EPROM, proceed as follows:

1.Shut off and lock out power to the electronics package.

2.Open electronics enclosure.

3.On the Main PCB, Figure 2-10,locate and remove old EPROM.

4.Replace with new EPROM.

5.Close electronics enclosure and power up system.

EPROM

MAIN PCB

P0020

Figure 2-10. Main PCB (Model 218A) EPROM Replacement

Rosemount Analytical Inc. A Division of Emerson Process Management

Installation 2-11

Image 33
Contents Oxymitter DR Essential Instructions Hazardous Area Oxymitter DR Table of ContentsList of Tables List of IllustrationsDefinitions PrefaceHazardous Area Oxymitter DR Section I. Identification Section II. Physical Data Section V. Health Hazard Data Section VI. Reactivity Data Section VII. Spill or Leak Procedures Section IX. Special Precautions Hazardous Area Oxymitter DR Technical Support Hotline Section Description and Specifications Component Checklist of Typical System Package ContentsSystem Overview System Features System ConfigurationTypical System Installation Flame Arrestor Ceramic Diffusion Assembly Abrasive Shield AssemblyFlame Arrestor Snubber Diffusion Assembly Upgrading the Hazardous Area Oxymitter DRView a View B Specifications Code Sensing Probe Type Product MatrixPart Number Description Calibration ComponentsMechanical Installation Section InstallationInstallation Selecting LocationHazardous Area Oxymitter DR Installation Hazardous Area Oxymitter DR with Abrasive Shield Mounting Plate Outline Without Abrasive Shield With Abrasive ShieldOrienting the Optional Vee Deflector Electrical Installation Pneumatic Installation Connect O2 and Heater Thermocouple SignalsOverview System Setup132 Analog Electronics SetupG02 G04 Westinghouse Model 218A ELEC- Tronics SetupWestinghouse Model TC200 VERI- Trim Electronics Setup 11. Main PCB Model TC200 Eprom Replacement 12. Main PCB Model 132 Eprom Replacement Westinghouse Model 132 Digital Electronics Setup14. Heater Power Supply HPS 3000 Jumper Configuration World Class 3000 Intelligent Field Transmitter Setup15. DR Probe Wired to the ZA8C or a V8C Converter Yokogawa ZA8C and AV8C Converter Electronics SetupGeneral Section Startup and OperationHazardous Area Oxymitter DR Hazardous Area Oxymitter DR Replacement Section Maintenance and ServiceOverview CalibrationHazardous Area Oxymitter DR Exploded View Terminal Block Replacement InstallHeater Strut Replacement Entire Probe ReplacementCell Replacement Heater Strut Assembly22220028 Ceramic Diffusion Element Replacement Replacement ProcedureThis refers to the ceramic diffusion element only General Also, do not get any cement onto the flame arrestor element Ceramic Diffusion Element ReplacementTermination Housing Connections Termination Housing WiringHazardous Area Oxymitter DR Section Troubleshooting Probe TroubleshootingSymptom Check Fault Remedy Hazardous Area Oxymitter DR Equipment Return Section Return of MaterialHazardous Area Oxymitter DR Replacement Parts for Probe Section Replacement PartsCell Replacement Kit Section Optional Accessories BY-PASS PackagesO2 Calibration GAS Sequencer Hazardous Area Oxymitter DR Appendix A. Updating DR Oxymitter to Full Oxymitter Section AppendicesHazardous Area Oxymitter DR Figure A-1. Component Replacement Upgrade ProcedureFigure A-2. Terminal Block and Wiring Section Index Hazardous Area Oxymitter DR Warranty Hazardous Area Oxymitter DR Serial no Order no