Page 38

Instruction Manual

IB-106-340CDR Original Issue January, 2002

Hazardous Area Oxymitter DR

g.THE YOKOGAWA ZA8C AND AV8C CONVERTER ELECTRONICS SETUP

The Hazardous Area Oxymitter DR probe can be wired to work with the Yokogawa®

ZA8C and AV8C Converters. Connect the cabling from the ZA8C or AV8C terminal to the probe terminal in the junction box as shown in Figure 2-15.

 

 

 

 

ZA8C

 

 

 

 

 

 

 

 

OR

 

 

 

 

 

 

 

 

AV8C

 

 

 

 

 

 

 

TERMINAL

 

 

 

 

 

 

 

}

3+

 

 

 

 

 

 

CELL

 

 

 

 

 

 

 

 

4–

 

 

COLD

 

 

 

}

 

 

 

 

 

 

 

 

JUNCTION

 

 

 

5+

 

 

DEVICE

 

THERMOCOUPLE

 

 

 

+

 

 

 

6–

 

 

 

 

 

}

 

 

 

 

 

 

COLD

7+

 

 

 

 

 

 

 

 

 

 

 

 

 

JUNCTION

8–

 

 

 

 

 

 

 

 

 

 

 

 

32

GND

PROBE

16

 

 

 

 

 

 

 

 

 

SHIELD

 

 

 

 

 

HEATER

 

 

 

 

 

 

17

 

 

 

 

 

 

 

 

 

 

 

 

 

 

+

 

+

 

ZA8C OR AV8C

 

 

-

-

 

 

 

 

 

 

 

 

ZA8C

OR

AV8C

CONVERTER

HEATER AND SIGNAL CABLE

SHIELD

STACK WALL

EXISTING

MOUNTING

OXYMITTER DR PROBE

WITH ADAPTER FLANGE

NOTES: 1. HEATER TEMPERATURE SET TO 1380oF (750oC)

2.THE GREATER MASS OF THE OXYMITTER DIRECT REPLACEMENT PROBE REQUIRES LONGER TIME TO HEAT UP. UPON STARTUP, THE YOKOGAWA ELECTRONICS MAY INDICATE AN ERROR BECAUSE THE PROBE HAS NOT REACHED TEMPERATURE SETPOINT IN THE NORMAL TIME. REMOVE POWER FROM THE YOKOGAWA ELECTRONICS OR PROBE

MODULE TO CLEAR THE ERROR, AND RESTORE POWER. THIS PROCEDURE MAY HAVE TO

 

BE REPEATED A COUPLE OF TIMES BEFORE PROBE OPERATING TEMPERATURE IS REACHED.

36210009

Figure 2-15. DR Probe Wired to the ZA8C or A V8C Converter

2-16 Installation

Rosemount Analytical Inc. A Division of Emerson Process Management

Image 38
Contents Oxymitter DR Essential Instructions Table of Contents Hazardous Area Oxymitter DRList of Illustrations List of TablesPreface DefinitionsHazardous Area Oxymitter DR Section I. Identification Section II. Physical Data Section V. Health Hazard Data Section VI. Reactivity Data Section VII. Spill or Leak Procedures Section IX. Special Precautions Hazardous Area Oxymitter DR Technical Support Hotline System Overview Section Description and SpecificationsComponent Checklist of Typical System Package Contents System Configuration System FeaturesTypical System Installation Flame Arrestor Snubber Diffusion Assembly Abrasive Shield AssemblyFlame Arrestor Ceramic Diffusion Assembly Upgrading the Hazardous Area Oxymitter DRView a View B Specifications Product Matrix Code Sensing Probe TypeCalibration Components Part Number DescriptionInstallation Section InstallationMechanical Installation Selecting LocationHazardous Area Oxymitter DR Installation Hazardous Area Oxymitter DR with Abrasive Shield Mounting Plate Outline With Abrasive Shield Without Abrasive ShieldOrienting the Optional Vee Deflector Electrical Installation Connect O2 and Heater Thermocouple Signals Pneumatic InstallationSystem Setup OverviewSetup 132 Analog ElectronicsWestinghouse Model 218A ELEC- Tronics Setup G02 G04Westinghouse Model TC200 VERI- Trim Electronics Setup 11. Main PCB Model TC200 Eprom Replacement Westinghouse Model 132 Digital Electronics Setup 12. Main PCB Model 132 Eprom ReplacementWorld Class 3000 Intelligent Field Transmitter Setup 14. Heater Power Supply HPS 3000 Jumper ConfigurationYokogawa ZA8C and AV8C Converter Electronics Setup 15. DR Probe Wired to the ZA8C or a V8C ConverterSection Startup and Operation GeneralHazardous Area Oxymitter DR Overview Section Maintenance and ServiceHazardous Area Oxymitter DR Replacement CalibrationHazardous Area Oxymitter DR Exploded View Install Terminal Block ReplacementEntire Probe Replacement Heater Strut ReplacementHeater Strut Assembly Cell Replacement22220028 This refers to the ceramic diffusion element only General Ceramic Diffusion Element ReplacementReplacement Procedure Ceramic Diffusion Element Replacement Also, do not get any cement onto the flame arrestor elementTermination Housing Wiring Termination Housing ConnectionsHazardous Area Oxymitter DR Symptom Check Fault Remedy Section TroubleshootingProbe Troubleshooting Hazardous Area Oxymitter DR Section Return of Material Equipment ReturnHazardous Area Oxymitter DR Section Replacement Parts Replacement Parts for ProbeCell Replacement Kit O2 Calibration GAS Sequencer Section Optional AccessoriesBY-PASS Packages Hazardous Area Oxymitter DR Section Appendices Appendix A. Updating DR Oxymitter to Full OxymitterHazardous Area Oxymitter DR Upgrade Procedure Figure A-1. Component ReplacementFigure A-2. Terminal Block and Wiring Section Index Hazardous Area Oxymitter DR Warranty Hazardous Area Oxymitter DR Serial no Order no