d.WESTINGHOUSE MODEL TC200 VERI- TRIM ELECTRONICS SETUP
Before beginning operation, it is important that the probe heater setpoint of the existing electronics be changed to support the Haz- ardous Area Oxymitter DR
1.EPROM replacement. To replace the EPROM, proceed as follows:
(a)Shut off and lock out power to the electronics package.
(b)Open electronics enclosure.
(c)On the main PCB, Figure
(d)Replace with new EPROMs (part number 1M03154G02) being careful to install U11 and U12 in their proper locations.
(e)Close electronics enclosure and power up system.
2.Heater Setpoint Adjustment. The ad- justment procedure required for the Model TC200 Electronics Package is as follows:
(a)Open keylocked enclosure to ac- cess membrane keyboard.
(b)Put controller in PAR (parameter) mode by depressing "LOCK" "▲" "%O2" "INC" "ACK" in sequence.
(c)Depress "ACK" pushbutton to clear display.
(d)Press "NUM" pushbutton.
(e)Using "INC", "DEC" buttons, dis- play parameter 125.
(f)Press "VAL" button.
(g)Using "INC", "DEC" buttons, change parameter 125 value to 15.4.
(h)Press "ENT" to save new value.