Hazardous Area Oxymitter DR
Instruction Manual
f.WORLD CLASS 3000 INTELLIGENT FIELD TRANSMITTER SETUP
The DR probe operates with a 115 VAC heater. Ensure that the voltage selection jumpers in the IFT or HPS, if used, are set
properly. Refer to Figure
configuration information, refer to
JUMPER
CONFIGURATION
ALWAYS DISCONNECT LINE VOLTAGE FROM INTELLIGENT FIELD TRANSMITTER BEFORE CHANGING JUMPERS.
LINE VOLTAGE | JUMPER |
SELECTION | (INSTALL) |
|
|
100 V.A.C. | JM3, JM7, JM2 |
|
|
120 V.A.C. | JM8, JM7, JM1 |
|
|
220 V.A.C. | JM6, JM5, JM2 |
|
|
240 V.A.C. | JM6, JM5, JM1 |
|
|
PROBE HEATER | JUMPER |
VOLTAGE SELECTION | (INSTALL) |
|
|
WORLD CLASS PROBE (44V) | JM10 |
|
|
218 PROBE (115V) | JM9 |
|
|
WORLD CLASS "DIRECT | JM9 |
REPLACEMENT" PROBE (115V) |
|
OR OXYMITTER DIRECT |
|
REPLACEMENT PROBE (115V) |
|
|
|
36210015
Figure 2-13. IFT 3000 Power Supply Board Jumper Configuration
If incorrect heater voltage is selected, damage to the Oxymitter DR may occur. Refer to IB-106-300NFX for additional configuration information. Always update the relevant labeling to reflect the set voltage.
|
|
|
|
|
|
| ALWAYS DISCONNECT LINE VOLTAGE |
| |||
|
|
|
|
|
|
| FROM HEATER POWER SUPPLY AND |
| |||
|
|
|
|
|
|
| |||||
|
|
| JUMPER |
| ANALOG ELECTRONICS (IF USED) |
| |||||
|
|
| CONFIGURATIONS | BEFORE CHANGING JUMPERS. |
| ||||||
|
|
|
|
|
|
|
|
|
|
|
|
| LINE VOLTAGE | JUMPER |
|
|
| HEATER |
| JUMPER |
|
| |
| SELECTION | (INSTALL) |
|
|
| POWER |
|
|
| ||
|
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
1 | 100/120 V.A.C. | JM4, JM1 |
|
|
| REMOTE |
| REMOVE JM2 |
|
| |
| 220/240 V.A.C. | JM5 |
|
|
| ON |
| *INSTALL JM2 |
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| PROBE HEATER | JUMPER |
|
| ELECTRONICS |
| JUMPER |
|
| ||
| VOLTAGE SELECTION | (INSTALL) |
|
|
| SELECTION |
|
|
| ||
|
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
| WORLD CLASS PROBE | JM7 |
|
| *ANALOG (EXISTING) |
| INSTALL JM3, JM6 |
|
| ||
| (44V) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| DIGITAL |
| REMOVE JM3, JM6 |
|
| |
|
|
|
|
|
|
|
|
|
| ||
| 218 PROBE (115V) | JM8 |
|
| (NEXT GENERATION) |
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
| DIRECT REPLACEMENT | JM8 |
|
|
|
|
|
|
|
| |
| WORLD CLASS OR |
|
|
|
|
|
|
|
|
|
|
| DR OXYMITTER |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| NOTES: |
|
|
|
|
|
|
|
|
|
|
| 1 100 V.A.C. OPERATION REQUIRES TRANSFORMER PART |
|
|
| |||||||
| NUMBER 1M02961G02. |
|
|
| 36210014 |
Figure 2-14. Heater Power Supply (HPS 3000) Jumper Configuration
Rosemount Analytical Inc. A Division of Emerson Process Management | Installation |