Emerson oxymitterdr hazardous area in-siu oxygen probe Ceramic Diffusion Element Replacement

Page 47

Hazardous Area Oxymitter DR

Instruction Manual

IB-106-340CDR Original Issue January, 2002

g.If the contact assembly is damaged, replace the strut or the contact pad. Instructions for replacing the contact pad are in the cell re- placement kit.

h.Remove and discard the corrugated seal. Clean the mating faces of the probe tube and retainer. Remove burrs and raised surfaces with a block of wood and crocus cloth. Clean the threads on the retainer and hub.

i.Rub a small amount of anti-seize compound on both sides of the new corrugated seal.

j.Assemble the cell and flange assembly, cor- rugated seal, and probe tube. Make sure the calibration tube lines up with the cali- bration gas passage in each component. Apply a small amount of anti-seize com- pound to the screw threads and use the screws to secure assembly. Torque to

4 N·m (35 in-lbs).

k.Install the termination housing per the instructions in paragraph 4-5steps c through e.

l.Apply anti-seize compound to the threads of the cell assembly, hub, and setscrews. Re- install the hub on the cell assembly. Using pin spanner wrenches, torque to 14 N·m

(10 ft-lbs). If applicable, reinstall the vee de- flector, orienting apex toward gas flow. Se- cure with the setscrews and anti-seize compound. Torque to 2.8 N·m (25 in-lbs).

m.On systems equipped with an abrasive shield, install the dust seal gaskets, with joints 180° apart.

n.Reinstall the probe and gasket on the stack flange.

o.Follow the instructions in paragraph 4-3bto install the Hazardous Area Oxymitter DR into the stack or duct. If there is an abrasive shield in the stack, make sure the dust seal gaskets are in place as they enter the 15° reducing cone.

p.Turn on power and monitor thermocouple output. It should stabilize at 29.3+0.2 mV.

Set reference air flow at 56.6 l/hr (2 scfh). After the Hazardous Area Oxymitter DR sta- bilizes, calibrate the unit. If new components have been installed, repeat calibration after 24 hours of operation.

4-8 CERAMIC DIFFUSION ELEMENT REPLACEMENT

NOTE

This refers to the ceramic diffusion element only.

a.

General

 

 

The diffusion element protects the cell from

4

 

particles in process gases. Normally, it does

 

 

not need to be replaced because the vee

 

 

deflector protects it from particulate erosion.

 

 

 

 

In severe environments, the filter may be

 

 

broken or subject to excessive erosion. Ex-

 

 

amine the ceramic diffusion element when-

 

 

ever removing the probe for any purpose.

 

 

Replace if damaged.

 

 

Damage to the ceramic diffusion element

 

 

may become apparent during calibration.

 

 

Compare probe response with previous re-

 

 

sponse. A broken diffusion element will

 

 

cause a slower response to calibration gas.

 

 

Hex wrenches needed to remove setscrews

 

 

and socket head screws in the following

 

 

procedure are available as part of a Probe

 

 

Disassembly Kit, Table 7-1.

 

b.

Replacement Procedure

 

 

1. Follow the instructions in paragraph

 

 

4-3ato remove the Hazardous Area

 

 

Oxymitter DR from the stack or duct.

 

2. Loosen setscrews, Figure 4-5,using hex wrench from Probe Disassembly Kit, Table 7-1,and remove vee deflec- tor. Inspect setscrews. If damaged, re- place with stainless setscrews coated with anti-seize compound.

Rosemount Analytical Inc. A Division of Emerson Process Management

Maintenance and Service 4-7

Image 47
Contents Oxymitter DR Essential Instructions Hazardous Area Oxymitter DR Table of ContentsList of Tables List of IllustrationsDefinitions PrefaceHazardous Area Oxymitter DR Section I. Identification Section II. Physical Data Section V. Health Hazard Data Section VI. Reactivity Data Section VII. Spill or Leak Procedures Section IX. Special Precautions Hazardous Area Oxymitter DR Technical Support Hotline System Overview Section Description and SpecificationsComponent Checklist of Typical System Package Contents System Features System ConfigurationTypical System Installation Upgrading the Hazardous Area Oxymitter DR Abrasive Shield AssemblyFlame Arrestor Ceramic Diffusion Assembly Flame Arrestor Snubber Diffusion AssemblyView a View B Specifications Code Sensing Probe Type Product MatrixPart Number Description Calibration ComponentsSelecting Location Section InstallationMechanical Installation InstallationHazardous Area Oxymitter DR Installation Hazardous Area Oxymitter DR with Abrasive Shield Mounting Plate Outline Without Abrasive Shield With Abrasive ShieldOrienting the Optional Vee Deflector Electrical Installation Pneumatic Installation Connect O2 and Heater Thermocouple SignalsOverview System Setup132 Analog Electronics SetupG02 G04 Westinghouse Model 218A ELEC- Tronics SetupWestinghouse Model TC200 VERI- Trim Electronics Setup 11. Main PCB Model TC200 Eprom Replacement 12. Main PCB Model 132 Eprom Replacement Westinghouse Model 132 Digital Electronics Setup14. Heater Power Supply HPS 3000 Jumper Configuration World Class 3000 Intelligent Field Transmitter Setup15. DR Probe Wired to the ZA8C or a V8C Converter Yokogawa ZA8C and AV8C Converter Electronics SetupGeneral Section Startup and OperationHazardous Area Oxymitter DR Calibration Section Maintenance and ServiceHazardous Area Oxymitter DR Replacement OverviewHazardous Area Oxymitter DR Exploded View Terminal Block Replacement InstallHeater Strut Replacement Entire Probe ReplacementCell Replacement Heater Strut Assembly22220028 This refers to the ceramic diffusion element only General Ceramic Diffusion Element ReplacementReplacement Procedure Also, do not get any cement onto the flame arrestor element Ceramic Diffusion Element ReplacementTermination Housing Connections Termination Housing WiringHazardous Area Oxymitter DR Symptom Check Fault Remedy Section TroubleshootingProbe Troubleshooting Hazardous Area Oxymitter DR Equipment Return Section Return of MaterialHazardous Area Oxymitter DR Replacement Parts for Probe Section Replacement PartsCell Replacement Kit O2 Calibration GAS Sequencer Section Optional AccessoriesBY-PASS Packages Hazardous Area Oxymitter DR Appendix A. Updating DR Oxymitter to Full Oxymitter Section AppendicesHazardous Area Oxymitter DR Figure A-1. Component Replacement Upgrade ProcedureFigure A-2. Terminal Block and Wiring Section Index Hazardous Area Oxymitter DR Warranty Hazardous Area Oxymitter DR Serial no Order no