Emerson ib-106-340cdr instruction manual Figure A-2. Terminal Block and Wiring

Page 62

Instruction Manual

Appendix A

January, 2002

Hazardous Area Oxymitter DR

i.In the left side of the termination housing, place the new termination designation la- bels over the labels on the existing terminal block. After placing the new labels, the ter- minal block should appear as shown in Figure A-2.

j.The existing wiring from the Oxymitter to the electronics may be reused. However, the wires will be carrying new signals as noted by the new labels. The 4-20 mA wires must be removed from the old electronics and re- terminated to the wires carrying the 4-20 mA O2 signal to the control room.

k.The wires carrying the heater power must be converted to carry AC power (90-250

AC TERMINAL

COVER

VAC, 50/60 Hz) for the Oxymitter. The re- terminations may be inside the old elec- tronics housing, which will function as a simple junction box. Alternatively, the old electronics may be removed and replaced with a smaller junction box.

l.Place the round error blink code and cali- bration instructions label on the inside of the right housing cover.

m.Install both housing covers.

n.Refer to the instruction bulletin provided with your upgrade kit, IB-106-340C, for startup and diagnostic information.

TERMINAL

BLOCK AC LINE

VOLTAGE PORT

LINE VOLTAGE

AC

L1

 

 

(90 TO 250 VAC)

AC

N

 

 

+

GROUND

 

LUGS

LOGIC I/O OR

-

 

CALIBRATION HANDSHAKE

+

 

4-20 mA

 

4-20

 

SIGNAL

 

-

 

 

 

SIGNAL

LEFT SIDE OF

 

PORT

OXYMITTER 4000

 

36210016

Figure A-2. Terminal Block and Wiring

A-2 Appendices

Rosemount Analytical Inc. A Division of Emerson Process Management

Image 62 Contents
Oxymitter DR Essential Instructions Table of Contents Hazardous Area Oxymitter DRList of Illustrations List of TablesPreface DefinitionsHazardous Area Oxymitter DR Section I. Identification Section II. Physical Data Section V. Health Hazard Data Section VI. Reactivity Data Section VII. Spill or Leak Procedures Section IX. Special Precautions Hazardous Area Oxymitter DR Technical Support Hotline System Overview Section Description and SpecificationsComponent Checklist of Typical System Package Contents System Configuration System FeaturesTypical System Installation Flame Arrestor Snubber Diffusion Assembly Abrasive Shield AssemblyFlame Arrestor Ceramic Diffusion Assembly Upgrading the Hazardous Area Oxymitter DRView a View B Specifications Product Matrix Code Sensing Probe TypeCalibration Components Part Number DescriptionInstallation Section InstallationMechanical Installation Selecting LocationHazardous Area Oxymitter DR Installation Hazardous Area Oxymitter DR with Abrasive Shield Mounting Plate Outline With Abrasive Shield Without Abrasive ShieldOrienting the Optional Vee Deflector Electrical Installation Connect O2 and Heater Thermocouple Signals Pneumatic InstallationSystem Setup OverviewSetup 132 Analog ElectronicsWestinghouse Model 218A ELEC- Tronics Setup G02 G04Westinghouse Model TC200 VERI- Trim Electronics Setup 11. Main PCB Model TC200 Eprom Replacement Westinghouse Model 132 Digital Electronics Setup 12. Main PCB Model 132 Eprom ReplacementWorld Class 3000 Intelligent Field Transmitter Setup 14. Heater Power Supply HPS 3000 Jumper ConfigurationYokogawa ZA8C and AV8C Converter Electronics Setup 15. DR Probe Wired to the ZA8C or a V8C ConverterSection Startup and Operation GeneralHazardous Area Oxymitter DR Overview Section Maintenance and ServiceHazardous Area Oxymitter DR Replacement CalibrationHazardous Area Oxymitter DR Exploded View Install Terminal Block ReplacementEntire Probe Replacement Heater Strut ReplacementHeater Strut Assembly Cell Replacement22220028 This refers to the ceramic diffusion element only General Ceramic Diffusion Element ReplacementReplacement Procedure Ceramic Diffusion Element Replacement Also, do not get any cement onto the flame arrestor elementTermination Housing Wiring Termination Housing ConnectionsHazardous Area Oxymitter DR Symptom Check Fault Remedy Section TroubleshootingProbe Troubleshooting Hazardous Area Oxymitter DR Section Return of Material Equipment ReturnHazardous Area Oxymitter DR Section Replacement Parts Replacement Parts for ProbeCell Replacement Kit O2 Calibration GAS Sequencer Section Optional AccessoriesBY-PASS Packages Hazardous Area Oxymitter DR Section Appendices Appendix A. Updating DR Oxymitter to Full OxymitterHazardous Area Oxymitter DR Upgrade Procedure Figure A-1. Component ReplacementFigure A-2. Terminal Block and Wiring Section Index Hazardous Area Oxymitter DR Warranty Hazardous Area Oxymitter DR Serial no Order no