11 Comparator Function
This function compares a display value with the upper limit (HI) and the lower limit (LO) and displays these results.
It is necessary to set the “comparator function (f6 0 ~ f6 7 )” parameters, upper limit value (HI) and lower limit value (LO) in the function table, in advance to use this function.
It is necessary to install option
Comparator Sign
Comparison results are displayed by indicators
Comparison Condition
weighing value < lower limit value | LO | is displayed and output. |
lower limit value < weighing value < upper limit value | OK | is displayed and output. |
upper limit value < weighing value | HI | is displayed and output. |
Parameter List and Word Definition
The “
The “digit” is a unit of display, and is equivalent to the minimum graduation. The “zero point” is the fundamental starting point to weighing.
Function table |
|
| Meaning and Purpose | |||||||||||
f6 | 0 | Having pressed the | SET | switch, the scale will always compare the | ||||||||||
|
| current display value. | ||||||||||||
f6 | 1 | The scale always compares the display value. | ||||||||||||
f6 | 2 | Pressing the |
|
| switch the scale always compares the display value | |||||||||
SET | ||||||||||||||
|
| if not | ||||||||||||
f6 | 3 | The scale always compares the display value if not | ||||||||||||
f6 | 4 | When the display value becomes stable after pressing the |
|
| ||||||||||
SET | ||||||||||||||
|
| switch, the scale compares the display value. If the |
|
| switch is | |||||||||
|
| SET | ||||||||||||
|
| pressed again the scale stops the comparison. | ||||||||||||
f6 | 5 | When the display value is stable the scale compares the display | ||||||||||||
|
| value. It does not compare on an unstable condition. | ||||||||||||
f6 | 6 | When the display value becomes stable while not | ||||||||||||
|
| pressing the |
| switch, the scale compares the display value. It | ||||||||||
|
| SET | ||||||||||||
|
| does not compare on an unstable condition. If the |
| switch is | ||||||||||
|
| SET | ||||||||||||
|
| pressed again the scale stops the comparison. | ||||||||||||
f6 | 7 | When the display value becomes stable and not | ||||||||||||
|
| compares the value. | ||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Caution |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
The upper limit value (HI) must be equal to or greater than the lower limit value (LO).
The parameters of the upper limit value (HI) and the final value (HI) use the same memory. The parameters of the lower limit value (LO) and the zero band (LO) use the same memory. The upper/lower comparator function, the simple batch function and the full/dribble batch function cannot be used at the same time because these parameters use common memory.