15.2

Parameter List

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Item

Display

Meaning and purpose

 

 

 

 

 

The type L scale is turned off after 5 minutes of no operation,

 

 

 

 

 

when displaying zero.

Automatic

f1 0

#

Not used

turning off

f1 1

 

Used

 

 

 

 

 

 

Transmission rate of the serial interface (RS-232C/ 422/485).

 

 

 

 

f4 0

#

2400bps

 

Baud rate

f4 1

 

4800bps

 

 

 

 

f4 2

 

9600bps

 

 

 

 

 

 

Mode selection for the serial interface (RS-232C/ 422/485).

 

 

 

 

f5 0

#

Stream mode. (Refer to “16.2 Stream Mode”)

 

 

 

 

f5 1

 

Command mode. (Refer to “16.3 Command Mode”)

 

 

 

 

f5 2

 

A data is output, when the

 

 

switch is pressed.

 

 

 

 

 

PRINT

 

 

 

 

 

f5 3

 

Auto-print +

 

 

 

 

 

 

 

 

When the display becomes a positive stable value above

 

 

 

 

 

 

near-zero, the scale outputs the data automatically. Next

Output mode

 

 

output can be performed after the display becomes near-

 

 

 

 

 

 

zero or a negative value.

 

 

 

 

f5 4`

Auto-print +/-

 

 

 

 

 

 

When the display becomes a stable value outside near-zero,

 

 

 

 

 

 

the scale outputs the data automatically. Next output can be

 

 

 

 

 

 

performed after the display becomes near-zero.

 

 

 

 

f5 5

 

At each finish of full/dribble batch function, the data is output.

 

 

 

 

 

 

Mode selection for the comparator

 

 

 

 

f6 0 #

Pressing the

 

 

 

 

switch, the scale always compares the

 

 

 

 

 

SET

 

 

 

 

 

 

 

current display

value.

 

 

 

 

f6 1

 

The scale always compares the display value.

Comparator

f6 2

 

Pressing the

 

 

 

 

switch, the scale always compares the

 

SET

 

 

 

 

 

 

 

display value

when not near-zero.

 

 

 

 

f6 3

 

The scale always compares the display value when not

 

 

 

 

 

 

near-zero.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Notes

The “#” shows factory settings.

The “near-zero” is within ±4 digits from zero point in the unit of kg.