14.2.3 Calibration of the Zero Point

Step 8 Press and hold the CAL switch to enter the calibration mode after

displayingnormal weighing for 30 minutes.

Then the Cal 0 is displayed.

Step 9 With nothing on the pan, press the ENTER switch while the stable mark is displayed. The scale stores the current condition as the zero point. .

Step 10 The scale displays 5pn1 for several seconds. If finishing the calibration mode at this stage (Zero calibration only), press CAL to proceed to step 13.

14.2.4 Span Calibration

Step 11 Set the value of the calibration mass using the following switches. (The initial value depends upon the particular model.)

^switch Selecting the value of a digit.

< switch Selecting a digit.

Step 12 Place the mass’ selected at step 11’ onto the pan. When the stable mark is displayed press the ENTER switch.

The scale then calculates the span and stores the value.

Step 13 The scale displays end at the finish.

Remove the mass from the pan.

Step 14 Press the CAL switch to return to the normal weighing mode.