Step 6

Press the

MODE

switch to display the

 

 

HI, blinking. (Final value).

Step

7

Set the final value using the following switches.

 

 

^ switch

Selecting the value of a digit.

 

 

< switch

Selecting a digit.

Step

8

Press the

 

 

switch to store the new

ENTER

 

 

parameter and display the OK, blinking,

 

 

(Preliminary value).

Step

9

Set the preliminary value using the following

 

 

switches.

 

 

 

 

^

<

switch

Selecting the value of a digit.

switch

Selecting a digit.

Step 10 Press the ENTER switch to store the new parameter and display the LO, blinking, (Zero band).

Step 11 Set the zero band using the following switches.

^ switch

Selecting the value of a digit

< switch

Selecting a digit.

Step 12 Press the ENTER switch to store the new parameter and display the weighing mode.

13.2 Operation and Performance (Examples)

Step

1

Select the parameter f6 8 of the function table.

Step

2

Set the parameters of the final value, preliminary value and zero band.

Step

3

The comparison result is always displayed.