Step

6

If one of

f6 0

,

f6 2

,

f6 4

,

f6 6

has

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

been selected, press the

SET

 

switch to

 

 

use the comparator.

Step

7

Press the

 

 

 

 

switch to display the

 

MODE

 

 

HI, blinking.

Step

8

Set the upper limit value by using the

 

 

following switches.

 

 

^ switch

 

Selecting the value of a digit.

 

 

< switch

 

Selecting a digit.

 

 

F switch

 

Selecting the polarity (+,-).

Step

9

Press the

 

 

 

switch to store the new

ENTER

 

 

parameter and display the blinking LO.

Step 10

Set the lower limit value by using the

 

 

following switches.

 

 

^ switch

 

Selecting the value of a digit.

 

 

< switch

 

Selecting a digit.

 

 

F switch

 

Selecting the polarity (+,-).

Step 11

Press the

 

 

 

 

switch to store the

ENTER

 

 

new parameters and display the

 

 

weighing mode.