2-7-3-3 Examples of Programming the Monitor

Note: The procedure is the same for the TM-2420/TM-2421/TM-2430, except periods 5 and 6, which are available only with the TM-2430.

(A) Example: Programming the Monitor for One (1) Period

Suppose that the start time of the monitoring session is 09:00 and you would like the monitor to measure the patient's blood pressure every 30 minutes. Remember that the Time for Period One (1) also defines the End of the monitoring session. The end time for Period 1 will default to the Time for Period One (1), in this case 09:00.

PERIOD

TIME

INTERVAL

DURATION

1

09:00-09:00

30 minutes

24 hours

 

 

Total:

24 hours

The following is the recorder setup according to these instructions.

Note: Set Period One (1) only. Leave the other Periods as “-”.

PERIOD 1

2

3

4

5

6

START INTERVAL

9

 

 

30 minutes

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

-

 

 

-

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

-

 

 

-

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

-

 

-

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

-

 

-

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

-

 

 

-

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 23: Monitor programming example 1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

37

Page 38
Image 38
A&D TM-2430-13 instruction manual Period Start Interval