ERROR CODE

POSSIBLE CAUSE

 

CORRECTION

 

 

 

 

E07

Manual stop

 

 

The

measurement

was

manually

 

 

 

 

 

interrupted by depression of the red

 

 

 

 

 

Start/Stop button.

 

 

 

E08

Korotkoff

 

sounds

are

Check that the lower microphone is

 

undetectable.

 

 

affixed with the sensing side (skin side)

 

 

 

 

 

facing the patient’s arm.

 

 

 

 

 

 

 

 

Note: The microphone should always

 

 

 

 

 

be taped to the arm with the supplied

 

 

 

 

 

affixing tape. Place "X" of tape over the

 

 

 

 

 

top of the microphone.

 

 

 

E11

More than 256 Korotkoff sounds

Check for a break in the microphone

 

are measured.

 

 

lead. Check for external damage to the

 

 

 

 

 

microphone. Microphone circuit may be

 

 

 

 

 

damaged.

 

 

 

 

 

 

 

 

 

 

Note: If a break in the microphone lead

 

 

 

 

 

or external damage has occurred, the

 

 

 

 

 

microphone

assembly

must

be

 

 

 

 

 

replaced.

 

 

 

 

 

E20

Pulse Error

 

 

Check the position of the lower

 

Pulse less than 35 beats or more

microphone and then re-correlate it to

 

than 200 beats per minute.

 

the subject.

 

 

 

 

 

 

 

 

 

Note: Certain physiological conditions

 

 

 

 

 

or

pharmaceutical treatment

may

 

 

 

 

 

excessively reduce or increase a pulse

 

 

 

 

 

rate. In cases of excessively low pulse

 

 

 

 

 

rates, have the subject walk or stand a

 

 

 

 

 

few minutes to raise the pulse rate

 

 

 

 

 

above 35 beats per minute.

 

 

E21

Diastolic greater than 160 mmHg

Outside

of

normal

physiological

 

or less than 40 mmHg.

 

parameters.

 

 

 

 

E22

Systolic greater than 280 mmHg

Outside

of

normal

physiological

 

or less than 60 mmHg.

 

parameters.

 

 

 

 

E23

Difference between systolic blood

Outside

of

normal

physiological

 

pressure

and

diastolic

blood

parameters.

 

 

 

 

 

pressure less than 10 mmHg.

 

 

 

 

 

 

 

E24

Difference between systolic blood

Confirm that the subject is not moving

 

pressure

and

diastolic

blood

during measurement.

Reposition

the

 

pressure

greater than

150

lower microphone and then correlate it

 

mmHg.

 

 

 

to the subject.

 

 

 

 

Note: If a measurement error occurs (other than E08) when the “Auto” mode is set, the monitor will automatically take a 2nd measurement. If a second measurement error occurs, a 3rd measurement will not be taken. The monitor will return to its programmed interval.