3-1-2 TM-2430 Series Error Codes

ERROR CODE

POSSIBLE CAUSE

 

CORRECTION

E00

Clock parameters have not been

Enter clock parameters.

 

set.

 

 

 

E03

An error code is displayed

Exhaust the air from the cuff completely.

 

without cuff inflation.

 

 

E04

Measurement is stopped due to a

Replace with new batteries. Auto mode

 

low battery.

 

 

is quit. Restart the Auto mode if you use

 

 

 

 

it.

E05

Inflation pressure does not reach

Wrap the cuff and connect to the

 

the target pressure.

 

monitor. If the error persists, there may

 

 

 

 

be an air leak and repair is necessary.

E06

Pressurization greater

than 320

Do not move and try to relax during the

 

mmHg.

 

 

measurement. If you can not clear the

 

 

 

 

error, repair is necessary.

 

 

 

E07

Controlled stop using STOP key

Do not press the STOP key when

 

 

 

 

unnecessary.

 

 

 

E08, E10

Pulse can not be detected.

Do not move and try to relax during the

 

 

 

 

measurement.

 

 

 

E20

Pulse less than 35 beats or more

Measure the blood pressure by other

 

than 200 beats per minute.

methods.

 

 

 

E21

Diastolic greater than 160 mmHg

Measure the blood pressure by other

 

or less than 40 mmHg.

 

methods.

 

 

 

E22

Systolic greater than 280 mmHg

Measure the blood pressure by other

 

or less than 60 mmHg.

 

methods.

 

 

 

E23

DSD greater than 150 mmHg or

Measure the blood pressure by other

 

less than 10 mmHg.

 

methods.

 

(DSD = difference between

 

 

systolic blood pressure and

 

 

diastolic blood pressure)

 

 

 

 

E30, E31

Air leaks in the cuff assembly.

Repair is necessary.

 

 

 

 

 

E32

Clock error

 

 

If the error persists, repair is necessary.

 

 

 

 

E50

Pressure offset error

 

Repair is necessary.

 

 

 

 

E52

Memory error

 

Repair is necessary.

 

 

 

E53

Defective battery contact

Replace batteries correctly. If the error

 

 

 

 

persists, repair is necessary.

 

 

 

 

E55, E56, E57

Exhaust error

 

Relax and do not move during the

 

 

 

 

measurement. If this error occurs

 

 

 

 

frequently, repair is necessary.

 

 

 

 

E60

Interval setting error

 

Enter the interval parameters correctly.

 

 

 

 

E70, 71, 72, 73

RS-232C error

 

Re-connect the communication cable. If

 

 

 

 

the error persists, repair is necessary.

 

 

 

 

 

E74

Voltage

decreases

during

Replace batteries with new ones and

 

communication.

 

restart communication.

 

 

 

E75

Protocol error due to external

Re-connect the communication cable. If

 

equipment

 

 

the error persists, repair is necessary.

 

 

 

 

 

50

Page 51
Image 51
A&D TM-2430-13 instruction manual 2 TM-2430 Series Error Codes