♦Histograms
After Partial Analysis parameters are defined, the data specified within the parameters will appear in red overlaying the Full Analysis data.
2-4-2 BP Mode
The BP Mode is used to select which blood pressure method will be used as the primary and secondary choices for displaying the blood pressure measurement data. A&D monitors employ either Korotkoff, oscillometric or dual Korotkoff and oscillometric measurement methods. If the current data file contains data from a dual system, you may choose Korotkoff, oscillometric or both methods to analyze the data.
The measurement method(s) specified in the BP Mode will be the basis for the statistical and graphical analysis performed on the opened data file. The method noted with a one (1) is the primary method and the method noted with a two (2) is the secondary method. If a primary method measurement error occurred during the monitoring session or a measurement is manually excluded (see
Method Setting | Description |
Oscillometric only | This option sets the primary blood pressure choice to |
| oscillometric and disables the secondary choice. If the |
| oscillometric reading for any measurement is invalid or has |
| been excluded by the user in the BP Data List, the |
| measurement will be excluded from graphical displays and |
| statistical computations. |
|
|
1: Oscillometric, 2: Korotkoff | This option sets the primary blood pressure choice to |
| oscillometric and the secondary choice to Korotkoff. If an |
| oscillometric reading is invalid, oscillometric readings will |
| be substituted with Korotkoff readings. If both readings |
| are invalid for a measurement, all data from the |
| measurement is excluded. |
|
|
Korotkoff only | This option sets the primary blood pressure choice to |
| Korotkoff and disables the secondary choice. If the |
| Korotkoff reading for any measurement is invalid or has |
| been excluded by the user in the BP Data List, the |
| measurement will be excluded from graphical displays and |
| statistical computations. |
|
|
1: Korotkoff, 2: Oscillometric | This option sets the primary blood pressure choice to |
| Korotkoff and the secondary choice to oscillometric. If a |
| Korotkoff reading is invalid, Korotkoff readings will be |
| substituted with oscillometric readings. If both readings |
| are invalid for a measurement, all data from the |
| measurement is excluded. |
|
|
11