Histograms

After Partial Analysis parameters are defined, the data specified within the parameters will appear in red overlaying the Full Analysis data.

2-4-2 BP Mode

The BP Mode is used to select which blood pressure method will be used as the primary and secondary choices for displaying the blood pressure measurement data. A&D monitors employ either Korotkoff, oscillometric or dual Korotkoff and oscillometric measurement methods. If the current data file contains data from a dual system, you may choose Korotkoff, oscillometric or both methods to analyze the data.

The measurement method(s) specified in the BP Mode will be the basis for the statistical and graphical analysis performed on the opened data file. The method noted with a one (1) is the primary method and the method noted with a two (2) is the secondary method. If a primary method measurement error occurred during the monitoring session or a measurement is manually excluded (see 2-6-2-1Excluding/Including BP Measurements”), a valid measurement in the secondary method will be substituted for it with the appropriate BP Mode selection.

Method Setting

Description

Oscillometric only

This option sets the primary blood pressure choice to

 

oscillometric and disables the secondary choice. If the

 

oscillometric reading for any measurement is invalid or has

 

been excluded by the user in the BP Data List, the

 

measurement will be excluded from graphical displays and

 

statistical computations.

 

 

1: Oscillometric, 2: Korotkoff

This option sets the primary blood pressure choice to

 

oscillometric and the secondary choice to Korotkoff. If an

 

oscillometric reading is invalid, oscillometric readings will

 

be substituted with Korotkoff readings. If both readings

 

are invalid for a measurement, all data from the

 

measurement is excluded.

 

 

Korotkoff only

This option sets the primary blood pressure choice to

 

Korotkoff and disables the secondary choice. If the

 

Korotkoff reading for any measurement is invalid or has

 

been excluded by the user in the BP Data List, the

 

measurement will be excluded from graphical displays and

 

statistical computations.

 

 

1: Korotkoff, 2: Oscillometric

This option sets the primary blood pressure choice to

 

Korotkoff and the secondary choice to oscillometric. If a

 

Korotkoff reading is invalid, Korotkoff readings will be

 

substituted with oscillometric readings. If both readings

 

are invalid for a measurement, all data from the

 

measurement is excluded.

 

 