2-6-2 BP Data List

The BP Data List function displays individual blood pressure measurement within an open data file in tabular form. With the BP Data List, you may input measurement comments, exclude/include a measurement within the graphics and statistics or link a measurement to the beginning of the Trends Screen.

The following table defines each heading within the Blood Pressure Data List:

Item

 

 

Description

#

 

 

Measurement number

 

 

 

DATE/TIME

 

Date and time of the measurement

 

 

 

 

SYS

 

 

Systolic blood pressure

 

 

 

 

DIA

 

 

Diastolic blood pressure

 

 

 

 

MAP

 

 

Mean arterial blood pressure, obtained by MAP=DIA+(SYS-DIA)/3

 

 

 

 

PUL

 

 

Pulse

 

 

 

 

DP

 

 

Double Product, obtained by DP=SYSPUL/100

 

 

 

STATUS

 

The status line has 3 positions. When filled with a character other than “–”,

 

 

 

each denotes a specific occurrence or condition.

Position

1

Denotes the BP method used for the specific measurement. “K” refers to

BP

method,

Korotkoff method, “O” refers to oscillometric method.

O or K

 

 

Position

2

Indicates that data was measured when the monitor’s EVENT button was

E (Event)

 

activated.

Position

3

Denotes the specific error code that occurred during the monitoring session,

Error code

 

see 3-1 Measurement Errors”.

 

 

 

Exclude

 

When this column has an “X”, the measurement has been excluded from the

 

 

 

statistics and graphs.

 

 

 

Comments

 

Space provided for user to enter patient diagnostic notations up to 20

 

 

 

characters.

 

 

 

Vertical

scroll

The vertical scroll bar on the right hand side of the screen helps scroll through

bar

 

 

the measurements. This may also be performed with the up and down arrow

 

 

 

keys, as well as the mouse.

 

 

 

Save button

 

Saves any edits, i.e. comments, exclusions or inclusions.

 

 

 

Close button

 

Closes the screen.

 

 

 

Note: Edits (comments, exclusions or inclusions) will not be saved.

 

 

 

Help button

 

Opens the Blood Pressure Data List help screen.

 

 

 

 