3. TROUBLESHOOTING

3-1 Measurement Errors

The following tables list the error messages that may occur during a monitoring session and offer information on possible causes and corrections.

Measurement errors may be shown on the blood pressure data list display and printed text data. If an error occurs (indicating by asterisks), corrective measures must be taken to use this software.

3-1-1 TM-2420/TM-2421 Error Codes

ERROR CODE

POSSIBLE CAUSE

 

 

CORRECTION

E01

The microphone has

become

Remove and reconnect the microphone

 

disconnected or there has been

connector at the front of the monitor.

 

excessive body motion.

 

 

 

E02

Air leak

 

 

 

Check that the air hose plug is tightly

 

 

 

 

 

connected to the air connector located

 

 

 

 

 

at the front of the monitor.

 

 

 

 

 

Check the connection between the air

 

 

 

 

 

hose plugs and the air bladder.

 

 

 

 

 

Check the air bladder for leaks.

E04

Low Battery, “c”, shown in the

Plug the AC adapter in the monitor.

 

monitor display.

 

 

The “c” shown on the display will

 

The monitor will not take a

disappear after 30 minutes. Wait the full

 

measurement,

print

data,

or

60 minutes charging time to ensure that

 

download data until the monitor is

the battery is fully charged.

 

charged.

 

 

 

Note: The monitor will hold all stored

 

 

 

 

 

readings in memory for two weeks after

 

 

 

 

 

the appearance of the low battery

 

 

 

 

 

indicator.

E05

Measurement time exceeds 100

Check E01 and E02 procedures.

 

seconds.

 

 

 

Reposition the lower microphone and

 

(Inflation time error)

 

 

then correlate it to the subject.

 

 

 

 

 

Check that the air hose is not held down

 

 

 

 

 

by a constrictive strap or similar

 

 

 

 

 

restraint which may disrupt the air flow.

E06

Pressurization

greater

than

320

Confirm that the subject is not moving

 

mmHg

 

 

 

during the measurement.

 

 

 

 

 

Clear the memory.

 

 

 

 

 

Caution: All measurement data will

 

 

 

 

 

be erased.

 

 

 

 

 

After clearing the memory, take one

 

 

 

 

 

measurement. If the cuff pressurizes to

 

 

 

 

 

approximately 195 mmHg, the system is OK.

 

 

 

 

 

Reset the monitor by gently inserting a

 

 

 

 

 

paper clip in the hole located on the

 

 

 

 

 

bottom of the monitor.

 

 

 

 

 

Caution: All measurement data, as

 

 

 

 

 

well as programmed intervals and

 

 

 

 

 

conditions will be erased.

 

 

 

 

 

If the monitor stops and then continues

 

 

 

 

 

to inflate above 260 mmHg, return the

 

 

 

 

 

unit to A&D for service.

 

 

 

 

 

 