Measurement Operations & Settings Measurement Functions The SENSe Subsystem
110 Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
command: :SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:TERMination
syntax: :SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect :TERMination
description: For the currently selected source, start the calibration and save the calibration values for a
defined termination reference measurement. See
:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:REFLectan ce[l]” on page112 for infor-
mation on setting the return loss value of your ref erence reflector.
parameters: none
response: none
example sens1:ret:cal:coll:term
affects: All return loss modules
command: :SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:TERMination?
syntax: :SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:TERMi nation?
description: Que ries the T-value (termination calibr ation value) for the retu rn loss module
parameters: none
response: Termination calibration value as a float in dB
example sens1:ret:cal:term? +6.5000E+001
affects: All return loss modules
command: :SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:VALues?
syntax: :SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:VALues?
description: Retu rns the the current ca libration values
1. monitor diode reference power
2. return loss diode reference power
3. monitor diode parasitics power
4. return loss diode parasitics power.
parameters: Retur ns power valu es in W
response: none
example sens1:ret:cal:val
affects: All return loss modules