Measurement Operations & Settings Signal Conditioning
158 Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
command: :OUTPutn[:CHANnel[ m]]:CORRection:COLLection:ZERO:ALL
syntax: :OUTPutn[:CHANn el[m]]:CORRection:COLLection:ZERO:ALL
description: Zero all available powermeter ch annels in the mainframe.
parameters: none
response: none
example: OUTP1:CORR:COLL:ZERO:ALL
affects: Powermeter modules; attenuator modules with power control, and return loss modules.
command: :OUTPutn[:CHANnel[ m]]:CORRection:COLLection:ZERO?
syntax: :OUTPutn[:CHANn el[m]]:CORRection:COLLection:ZERO?
description: Queries the status of the last :OUTPutn[:CHANnel[ m]]:CORRection :COLLection:ZE RO opera-
tion.
parameters: none
response: integer 0 = OK, otherwise not OK.
example: OUTP:CORR:COLL:ZER0? 0<END>
affects: A ttenuator modul es with power c ontrol.