GPIB Error Strings Error Codes
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition 267
New -305 "Internal messaging erro r (StatCmdError)"
"Internal messaging error (StatCmdNotAllowed)"
"Internal messaging error (StatWrongLength)"
"Internal messaging error (StatWrongReceiver) "
"Internal messaging error (StatBufAllocError) "
"Internal messagin g error (StatD PRamFull)"; }
"Internal messagin g error (Sta tSemError)"
Description:
An error has occured in the instru ment communication sy stem. Please re-
port this error with a description of the circumstances that generated the
error and the config uration of the system.
New -306 "Channel doesn't s upport this command (Stat CmdUnknownFor Slave)"
Description:
Slave channels have limited functionality. The module supports this com-
mand, but the command must be sent to th e master channel.
New -307 "Channel without head connection (StatHeadless)"
Description:
The channel supports this command, but it cannot be executed because
the optical measurement head is not plugged into the interface module.
Standard -310 “System error”
[Indicates that some error, termed “s ystem error” by the de vice, has oc-
curred. This code is device-depe ndent.]
Standard -321 “Out of memory”
[An internal operation needed mo re memory than was av ailable.]
New -322 "Flash programming error (StatFlashEraseFailed)"
"Flash programming error (StatFla shWriteFailed)"
"Flash programming error (StatFlashDataCntError)"
"Flash programming error (StatFla shDPAlgoFailed)"
Description:
An error has occured in a module . Please report this err or with a descrip-
tion of the circumstances that generated the error and the config uration of
the system.
Table18 Overview for Supported Strings
New/Old/Standard
Error
Number String