Root Layer Command Measurement Operations & Settings
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition 81
command: :SLOT[n]:EMPTy?
syntax: :SLOT[n]:EMPTy?
description: Queries whether the m odule slot is empty.
parameters: none
response: A boolean value: 0: there is a module in the slot
1: the module slot is empty
examples: slot1:empt? 0<END> There is a module in s lot1
affects: Independent of module type
command: :SLOT[n]:IDN?
syntax: :SLOT[n]:IDN?
description: Returns information about the module.
parameters: none
response: HEWLETT-PACKARD:
mmmm:
ssssssss:
rrrrrrrrrr:
manufacturer
instrument model number (for examp le 81533B)
serial number
date of firmware revision
example: slot1:idn?
HEWLETT-PACKARD, 81533B,3411G06054,07-Aug-98<END>
NOTE The Agilent 81640A/80A/82A/89A Tunable Laser modules will always return
Agilent Technologies as the manufacturer.
All other Agilent 8163A Series modules return Agilent Technologies as the
manufacturer.
The HP 8153A Series modules will always return Agilent Technologies as the
manufacturer.
See *IDN? on page59 fo r informatio n on mainfram e identity st rings.
affects: Inde pendent of mo dule type
command: :SLOT[n]:OPTions?
syntax: :SLOT[n]:OPTions?
description: Returns information about a module’s options.
parameters: none
response: A string.
example: slot1:opt? NO CONNECTOR OPTION, NO INSTRUMENT OPTIONS<END>
affects: Independent of module type