Lambda Scan Applications The Agilent 816x VXIplug&play Instrument Driver
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition 235
Lambda Scan Applications
These functions combine multiple SCPI commands into a single, functional
operation. They are designed to allow q uick and easy access to common
instrument command sequences.
These application functions allow you to perform one of the following
applications:
A Lambda Scan - a Lambda Logging operation where an A gilent
8164A/B Lightwave Measurement System with a back-loa dable
Tunable Laser module and up to four Power Meters ins talled, performs
a wavelength sweep where the Tunable Laser module and Power
Meters are coordinated with each othe r.
A Multi Frame Lambda Scan - a Lambda Logging operation where an
Agilent 8164A/B Lightwave Measurement System with a back-loadable
Tunable Laser module performs a wavelength sweep and the Tunable
Laser modu le is coo rdinate d with Po wer Mete rs that a re insta lled in t he
mainframe and in other mainframes. Thes e mainframes must be
connected to the GPIB bus and have their Inp ut Trigger Connector
connected to the Output Trigger Connector of the Agilent 8164A/B
Lightwave Measurement System mainframe.
The following two functions apply to both Lambda Scan and Multi Frame
Lambda Scan applications:
The Set Lambda Scan Wavelength
(hp816x_set_LambdaScan_wavelength) function allows you to use a
different wavelength than 1550 nm during a Lambda Scan operation. All
Power Meters taking part in the Lambda Scan oper ation will be set to
the chosen wavelength.
The Enable High Sweep Speed (hp816x_enableHighSweepSp eed)
function enables/disables the highest available sweep speed (40
nanometer s per second ) for Lambda Sc an operatio ns. The Lamb da Scan
operation chooses the highest possible sweep speed for the chosen
step size.
If you choose Enable, the highest sweep speed possible will be used.
This may lead to less accurate meas urements.
If you choose Disable, the highest sweep sp eed will not be used.