Instrument Setup and Statu s Status Reporting The STATus Subsystem
66 Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
command: :STATus:OPERation:ENABle [:LEVel0]
syntax: :STATus:OPERation:ENABle[:LEVel0]<wsp><value>
description: Sets the bits in the Operational Status Enable Su mmary Mask (OSESM) that enable the con-
tents of the OSESR to affect the Status Byte (STB).
Setting a bit in this register to 1 enables the corresponding bit in the OSE SR to affect bit 7 of
the Status Byte.
parameters: The bit value for the OSESM as a 16-bit signed integer val ue (0 .. +32767)
The default value is 0.
response: none
example: stat:oper:enab 128
command: :STATus:OPERation:ENABle [:LEVel0]?
syntax: :STATus:OPERation:ENABle[:LEVel0]?
description: Returns the OSESM for the OSE SR
parameters: none
response: The bit value for the operation enable m ask as a 16-bit signed int eger value (0 .. +32767)
example: stat:oper:enab? +128 <END>
command: :STATus:OPERation[:EVENt]:LEVel1?
syntax: :STATus:OPERation[:EVENt]:LEVel1?
description: Returns the Operational Status Eve nt Summary Register (OSESR) for slots 15 to 17 of the
Agilent8166A/B Lightwave Multichannel System.
parameters: none
response: The sum of the results for the slots (a 16-bit signed integer value, where 0 value 32767):
Bits Mnemonics Decimal Value
Agilent 8166A/B
15-4 Not used 0
3Slot 17 Summary 8
2Slot 16 Summary 4
1Slot 15 Summary 2
0Not used 0
example: stat:oper:leve l1? +0<END>