Status Reporting The STATus Subsystem Instrument Setup and Status
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition 65
command: :STATus:OPERation:CONDition[:LEVel0]?
syntax: :STATus:OPERation:CONDition[:LEVel0]?
description: Reads the Operational Status Condition Summary Register.
parameters: none
response: The sum of the results for the individual slo ts (a 16-bit signed integer value, where 0 value
32767):
Bits Mnemonics Decimal Value
Agilent 8163A/B Agilent 8164A/B Agilent 8166A/B
15 Not used Not used Not used 0
14 No t used Not used Slot 14 Summary 1 6384
13 No t used Not used Slot 13 Summary 8 192
12 No t used Not used Slot 12 Summary 4 096
11 No t used Not used Slot 11 Summary 2 048
10 No t used Not used Slot 10 Summary 1 024
9 Not used Not used Slot 9 Summary 512
8 Not used Not used Slot 8 Summary 256
7 Not used Not used Slot 7 Summary 128
6 Not used Not used Slot 6 Summary 64
5 Not used Not used Slot 5 Summary 32
4Not used Slot 4 Summary Slot 4 Summary 16
3 Not used Slot 3 Summary Slot 3 Summary 8
2 Slot 2 Summary Slot 2 Summary Slot 2 Summary 4
1 Slot 1 Summary Slot 1 Summary Slot 1 Summary 2
0 Not used Slot 0 Summary Level 1 Summary 1
example: stat:oper:cond? +0<END>