Triggering - The TRIGger Subsystem

Measurement Operations & Settings

command:

:TRIGger[n][:CHANnel[m]]:OUTPut:REARm

 

syntax:

:TRIGger[n][:CHANnel[m]]:OUTPut:REARm<wsp>OFFON01

description:

Sets the arming response of a channel to an outgoing trigger.

N O T E

See Table Table 9 , for information on how this command affects the generation of

 

output triggers using power measurements.

parameters:

A boolean value:

OFF or 0: trigger rearming disabled

 

 

ON or 1: trigger rearming enabled (default)

N O T E

If you return to Local control, all modules return to the default setting.

response:

none

 

example:

trig1:outp:rearm 1

 

affects:

All Agilent 8163A/B series power meters, and Agilent 8161x series return loss modules.

dual sensors:

Can only be sent to master channel, slave channel is also affected.

 

 

 

command:

syntax:

description:

parameters:

response:

example:

affects:

dual sensors:

:TRIGger[n][:CHANnel[m]]:OUTPut:REARm?

:TRIGger[n][:CHANnel[m]]:OUTPut:REARm?

Returns the arming response of a channel to an outgoing trigger.

none

A boolean value: 0: trigger rearming disabled (default) 1: trigger rearming enabled.

trig1:outp:rearm? 0<END>

All Agilent 8163A/B series power meters, and Agilent 8161x series return loss modules.

Can only be sent to master channel, slave channel parameters are identical.

command:

syntax:

description:

parameters:

response:

example:

:TRIGger:CONFiguration

:TRIGger:CONFiguration<wsp><triggering mode>

Sets the hardware trigger configuration with regard to Output and Input Trigger Connectors.

0 or DISabled:

Trigger connectors are disabled.

1 or DEFault:

The Input Trigger Connector is activated, the incoming trigger response

 

for each slot “:TRIGger[n][:CHANnel[m]]:INPut” on page 173 deter-

 

mines how each slot responds to an incoming trigger, all slot events

 

(see “:TRIGger[n][:CHANnel[m]]:OUTPut” on page 176) can trigger the

 

Output Trigger Connector.

2 or PASSthrough:

The same as DEFault but a trigger at the Input Trigger Connector gener-

 

ates a trigger at the Output Trigger Connector automatically.

3 or LOOPback:

The same as DEFault but a trigger at the Output Trigger Connector gen-

 

erates a trigger at the Input Trigger Connector automatically.

none

 

trig:conf dis

 

Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition

177

Page 177
Image 177
Agilent Technologies 8166A, B, 8163A, 8164A 177, Output triggers using power measurements, TRIGger nCHANnel mOUTPutREARm

8163A, 8164A, 8166A, B specifications

Agilent Technologies B,86100A is a high-performance oscilloscope and signal integrity analyzer designed primarily for advanced digital communications applications. As a versatile tool, it supports a wide range of testing needs, making it indispensable for engineers and researchers involved in the development and testing of high-speed digital signals.

One of the standout features of the B,86100A is its capability to analyze signals with various bandwidths, accommodating both current and emerging communication standards. The device features a sampling rate of up to 80 GS/s and bandwidth capabilities of 33 GHz to ensure high accuracy in capturing fast signal transitions, which is critical for ensuring the integrity of complex digital signals.

The B,86100A employs Agilent's proprietary digital signal processing (DSP) technology, which significantly enhances measurement precision and reduces noise, enabling users to obtain clearer insights into signal behavior. Its advanced triggering capabilities allow for precise signal capture, making it particularly useful in troubleshooting and validating high-speed designs, as well as in evaluating the performance of optical and electrical devices.

In addition to its high-speed capabilities, the B,86100A offers a robust set of measurement tools including jitter analysis, eye diagram analysis, and equalization assessment. These features allow engineers to effectively analyze signal quality and address potential issues related to signaling distortions and inter-symbol interference.

The graphical user interface of the B,86100A is intuitive, enabling users to efficiently navigate through measurement options and visualize data results. Customizable measurement setups streamline workflow, ensuring that users can quickly adapt their tests to evolving project requirements.

Another key characteristic of the B,86100A is its modularity. The system supports a variety of plug-in modules, which can be tailored to specific application needs, such as different types of optical and electrical signals. This flexibility not only extends the operational capability of the instrument but also makes it a future-proof investment as technology continues to evolve.

In summary, Agilent Technologies B,86100A combines high-speed acquisition with advanced processing capabilities, making it an essential instrument for anyone involved in high-speed digital design and testing. With its ability to deliver precise measurements and extensive analysis features, it empowers engineers to achieve optimal performance and reliability in their systems.