Measurement Functions The SENSe Subsystem Measurement Operations & Settings
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition 103
command: :SENSe[n]:[CHANnel[m]]:POWer:RANGe[:UPPer]?
syntax: :SENSe[n]:[CHANnel[m]]:POWer:RANGe[:UPPer]?
description: Returns the range setting for the mo dule. For a return loss module, re turns the power range
of the return loss diode.
parameters: none
response: The range setting as a float value in dBm
(110 value +30).
example: sens1:pow:rang? -2.00000000E+001<END>
affects: All power meters and return loss modules.
dual sensors: Master and slave channels are in dependent.
command: :SENSe[n]:[CHANnel[m]]:POWer:RANGe:MONitor[:U PPer]
syntax: :SENSe[n]:[CHANnel[m]]:POWer:RANGe:MONitor[:UPPer]<wsp><value>[DBM]
description: Sets the power range for a retun loss module’s monitor diode.
The range changes at 10dBm intervals. The corresponding ranges for linear measurements
(measurements in Watts) is given below:
Range
+30dBm
+20dBm
+10dBm
0dBm
10dBm
20dBm
30dBm
40dBm
Upper Linear
Power Limit
1999.9 mW
199.99 mW
19.999mW
1999.9µW
199.99µW
19.999µW
1999.9 nW
199.99 nW
Range
50dBm
60dBm
70dBm
80dBm
90dBm
100d Bm
110d Bm
Upper Linear
Power Limit
19.999nW
1999.9 pW
199.99 pW
19.999pW
1.999 pW
0.199 pW
0.019 pW
parameters: The range as a float value in dBm. The number is rounded to the closest multiple of 10, be-
cause the range changes at 10dBm intervals. Units are in dBm.
response: none
example: sens1:pow:rang:mon -20DBM
affects: All return loss modules.
dual sensors: Master and slave channels are in dependent.