Error Codes GPIB Error Strings
262 Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
Old -211 “Trigger ignored”
Description:
A trigger has been detected but ignored because of timing contraints. (For
Example: average time to large).
Old -212 “Arm i g no r ed ”
Description:
The user can set the automatic re -arming option for inpu t and output trig-
ger events (see “Triggering - The TRIGger Subsystem” on page171). When
this error occurs, the device ignores the setting because the current mod-
ule status does not allow the change of trigger settings.
Old -213 “Init ignored”
Description:
The INIT:IMM command (page88) initiates a trigger and completes a full
measurement cycle. The continuous measurement must be DISABLED.
This error code is generated if the powermeter is still in cont. measure-
ment mode.
Old -220 "Parameter error (StatParmError)”
Description:
The user has passed a paramet er that cannot be chan ged in this way. The
device cannot detect one of the following more specific errors:
Old -220 -220, "Parameter error (StatParmOutOfRan ge)"
Description:
The user has passed a parameter that exceeds the valid range for this pa-
rameter.
Old -220 "Parameter error (StatParmIllegalVal)"
Description:
The user has passed a parameter that does not match a valu e in a list of
possible values.