Chapter 5 Service

Self-Test Procedures

632

P6V hardware test failed This test checks the the status of voltage and

 

current error amplifiers for +6V power circuit. If both amplifiers are not

 

operational, the power supply will beep and the error annunciator will be on.

633

P25V hardware test failed This test checks the the status of voltage and

 

current error amplifiers for +25V power circuit. If both amplifiers are not

 

operational, the power supply will beep and the error annunciator will be on.

634

N25V hardware test failed This test checks the the status of voltage and

 

current error amplifiers for -25V power circuit. If both amplifiers are not

 

operational, the power supply will beep and the error annunciator will be on.

5

101

Page 102
Image 102
Agilent Technologies E3631A manual 101