Chapter 3 Calibration Procedures
General Calibration/Adjustment Procedure
To calibrate the output voltages and currents of the power supply from the front panel, proceed as follows:
1Unsecure the power supply.
To calibrate the output voltage and current, you must unsecure the power supply according to the procedure given on page 62.
2Disconnect all loads from the power supply and connect a DVM across output terminals of the +6V output.
3Turn on the calibration mode.
cal mode
Turn on the calibration mode by pressing the "Calibrate" key while simultaneously turning on the power supply then continue to hold the "Calibrate" key for about 5 seconds until a beep is heard.
Make sure that the power supply is in "CV" mode. If the power supply is not in "CV" mode, an error occurs.
4Move down a level to the voltage calibration mode for the +6V supply.
cal setup 1
The display shows the above message to indicate that the power supply is ready for the voltage calibration of +6V supply.
5Select the first voltage calibration point for the +6V supply.
v lo +0.1000 v
The display shows the first voltage calibration point for the +6V supply.
6Read the DVM and change the first voltage value on the display to match the measured voltage.
For example, if the DVM reading is 0.0860 V, adjust the voltage to 0.0860 V using the knob and resolution selection keys.
v lo +0.0860 v