Chapter 3 Calibration Procedures
Constant Current (CC) Verifications
Normal Mode Current Noise (CC Ripple and Noise)
The normal mode current noise is specified as the rms output current in a frequency range 20 Hz to 20 MHz with the power supply in constant current operation.
1Turn off the power supply and connect the output to be tested as shown in Figure
2Turn on the power supply and select the output to be tested using the meter selection key on the front panel. Enable the outputs and set the display to the limit mode. When the display is in the limit mode, program the current to full- scale value and the voltage to the maximum programmable value.
3The output current should be at the
4Divide the reading on the rms voltmeter by the load resistance to obtain rms current. The readings should be within the limit specified below for each output tested.
Output | Specification |
|
|
+6V | 2 mA |
|
|
+25V | 0.5 mA |
|
|
0.5 mA | |
|
|
5Repeat steps (1) through (4) for the remaining outputs.
58