Chapter 3 Calibration Procedures

Constant Current (CC) Verifications

Normal Mode Current Noise (CC Ripple and Noise)

The normal mode current noise is specified as the rms output current in a frequency range 20 Hz to 20 MHz with the power supply in constant current operation.

1Turn off the power supply and connect the output to be tested as shown in Figure 3-1with a load resistor (1.29 for +6V supply and 259 for ±25V supplies) across output terminals to be tested. Connect a rms voltmeter across the load resistor. Use only a resistive load for this test.

2Turn on the power supply and select the output to be tested using the meter selection key on the front panel. Enable the outputs and set the display to the limit mode. When the display is in the limit mode, program the current to full- scale value and the voltage to the maximum programmable value.

3The output current should be at the full-scale rating with the CC annunciator on. If not lit, adjust the load so that the output voltage drops slightly until the CC annunciator lights.

4Divide the reading on the rms voltmeter by the load resistance to obtain rms current. The readings should be within the limit specified below for each output tested.

Output

Specification

 

 

+6V

2 mA

 

 

+25V

0.5 mA

 

 

-25V

0.5 mA

 

 

5Repeat steps (1) through (4) for the remaining outputs.

58

Page 59
Image 59
Agilent Technologies E3631A manual Normal Mode Current Noise CC Ripple and Noise