Chapter 2 Quick Start
To check the rated currents of the power supply
28
6 Ensure that the current can be adjusted from zero to the maximum
rated value. 1
Adjust the knob until the ammeter indicates 0 amps and then until the ammeter
indicates 5.0 amps.
7 Check the current function for the +25V supply.
Disable the outputs by pressing the key and connect a short
across (+) and (COM) output terminals of the ±25V supply with an insulated
test lead. Repeat steps (3) through (6) after selecting the meter and adjust
selection key for the +25V supply.
8 Check the current function for the -25V supply.
Disable the outputs by pressing the key and connect a short
across (-) and (COM) output terminals of the ±25V supply with an insulated
test lead. Repeat steps (3) through (6) after selecting the meter and adjust
selection key for the -25V supply.
Note If an error has been detected during the output checkout procedures, the
ERROR annunciator will turn on. See "Error Messages" for more
information, starting on page 113 in chapter 5 of the User's Guide.
1You can use the resolution selection keys to move the blinking digit to the
right or left when setting the current.
+25V
Output On/Off
-25V
Output On/Off