Chapter 3 Calibration Procedures Performance Test Record for Agilent E3631A

Performance Test Record for Agilent E3631A

CV Performance Test Record

Test Description

Actual

Specifications

 

 

Result

Upper Limit

Lower Limit

 

 

 

 

 

 

 

CV Programming Accuracy @ 0 volts

 

 

 

(DVM reading)

 

 

 

+6V supply

 

+0.0050 V

-0.0050 V

+25V supply

 

+0.0200 V

-0.0200 V

-25V supply

 

+0.0200 V

-0.0200 V

 

 

 

 

CV Readback Accuracy @ 0 volts

 

 

 

+6V supply

 

DVM + 0.0050 V

DVM - 0.0050 V

+25V supply

 

DVM + 0.0100 V

DVM - 0.0100 V

-25V supply

 

-(DVM + 0.0100) V

-(DVM - 0.0100) V

 

 

 

 

CV Programming Accuracy @ Full Scale

 

 

 

(DVM reading)

 

 

 

+6V supply

 

6.0110 V

5.9890 V

+25V supply

 

25.0325 V

24.9675 V

-25V supply

 

-25.0325 V

-24.9675 V

 

 

 

 

CV Readback Accuracy @ Full Scale

 

 

 

+6V supply

 

DVM + 0.0110 V

DVM - 0.0110 V

+25V supply

 

DVM + 0.0225 V

DVM - 0.0225 V

-25V supply

 

-(DVM + 0.0225) V

-(DVM - 0.0225) V

 

 

 

 

CV Load Regulation

 

Maximum change:

 

+6V supply

 

<2.6 mV

 

+25V supply

 

<4.5 mV

 

-25V supply

 

<4.5 mV

 

 

 

 

 

CV Line Regulation

 

Maximum change:

 

+6V supply

 

<2.6 mV

 

+25V supply

 

<4.5 mV

 

-25V supply

 

<4.5 mV

 

 

 

 

 

CV Ripple/Noise

 

 

 

+6V supply

 

<2 mV p-p, 0.35 mV rms for each of the

+25V supply

 

three outputs

 

-25V supply

 

 

 

 

 

 

 

CV Current Noise (Common Mode)

 

 

 

+6V supply

 

<1.5 mA for each of the three outputs

+25V supply

 

 

 

 

-25V supply

 

 

 

 

 

 

 

Load Transient Response Time

 

 

 

+6V supply

 

<50 msec for each of three outputs

+25V supply

 

 

 

 

-25V supply

 

 

 

 

 

 

 

3

59

Page 60
Image 60
Agilent Technologies manual Performance Test Record for Agilent E3631A, CV Performance Test Record