Chapter 3 Calibration Procedures Performance Test Record for Agilent E3631A
Performance Test Record for Agilent E3631A
CV Performance Test Record
Test Description | Actual | Specifications | ||
|
| |||
Result | Upper Limit | Lower Limit | ||
| ||||
|
| |||
|
|
|
| |
CV Programming Accuracy @ 0 volts |
|
|
| |
(DVM reading) |
|
|
| |
+6V supply |
| +0.0050 V | ||
+25V supply |
| +0.0200 V | ||
| +0.0200 V | |||
|
|
|
| |
CV Readback Accuracy @ 0 volts |
|
|
| |
+6V supply |
| DVM + 0.0050 V | DVM - 0.0050 V | |
+25V supply |
| DVM + 0.0100 V | DVM - 0.0100 V | |
| ||||
|
|
|
| |
CV Programming Accuracy @ Full Scale |
|
|
| |
(DVM reading) |
|
|
| |
+6V supply |
| 6.0110 V | 5.9890 V | |
+25V supply |
| 25.0325 V | 24.9675 V | |
| ||||
|
|
|
| |
CV Readback Accuracy @ Full Scale |
|
|
| |
+6V supply |
| DVM + 0.0110 V | DVM - 0.0110 V | |
+25V supply |
| DVM + 0.0225 V | DVM - 0.0225 V | |
| ||||
|
|
|
| |
CV Load Regulation |
| Maximum change: |
| |
+6V supply |
| <2.6 mV |
| |
+25V supply |
| <4.5 mV |
| |
| <4.5 mV |
| ||
|
|
|
| |
CV Line Regulation |
| Maximum change: |
| |
+6V supply |
| <2.6 mV |
| |
+25V supply |
| <4.5 mV |
| |
| <4.5 mV |
| ||
|
|
|
| |
CV Ripple/Noise |
|
|
| |
+6V supply |
| <2 mV | ||
+25V supply |
| three outputs |
| |
|
|
| ||
|
|
|
| |
CV Current Noise (Common Mode) |
|
|
| |
+6V supply |
| <1.5 mA for each of the three outputs | ||
+25V supply |
| |||
|
|
| ||
|
|
| ||
|
|
|
| |
Load Transient Response Time |
|
|
| |
+6V supply |
| <50 msec for each of three outputs | ||
+25V supply |
| |||
|
|
| ||
|
|
| ||
|
|
|
|
3
59