Chapter 1 Specifications

Performance Specifications

Programming Resolution

 

 

 

+6V Output

+25V Output

-25V Output

Voltage

0.5 mV

1.5 mV

1.5 mV

Current

0.5 mA

0.1 mA

0.1 mA

Readback Resolution

 

 

 

+6V Output

+25V Output

-25V Output

Voltage

0.5 mV

1.5 mV

1.5 mV

Current

0.5 mA

0.1 mA

0.1 mA

Meter Resolution

 

 

 

+6V Output

+25V Output

-25V Output

Voltage

1 mV

10 mV

10 mV

Current

1 mA

1 mA

1 mA

Transient Response Time

Less than 50 msec for output recover to within 15 mV following a change in output current from full load to half load or vice versa

Command Processing Time

Programming Commands : Maximum time for output to change after receipt of APPLy

and SOURce commands : <50 msec

Readback Command : Maximum time to readback output by MEASure? command : <100 msec

The Other Commands : < 50 msec

Tracking Accuracy

The ±25V outputs track each other within ± (0.2% of output + 20 mV).