Chapter 3 Calibration Procedures
Constant Current (CC) Verifications
55
3
5Readback the output current from the selected output over the remote
interface by sending the command:
MEAS:CURR? {P6V|P25V|N25V}
6Record the value displayed on the controller. This value should be within the
limits specified below for each output tested.
7Program the selected output current to the full scale value (see Table 3-2) by
sending the commands:
CURR 5.0 for the +6V supply or
CURR 1.0 for the ±25V supply
8Divide the voltage drop (DVM reading) across the current monitoring resistor
(RM) by its resistance to convert to amps and record this value (IO). This value
should be within the limits specified below for each output tested.
9Readback the output current from the selected output over the remote
interface by sending the command:
MEAS:CURR? {P6V|P25V|N25V}
10 Record the value displayed on the controller. This value should be within the
limits specified below for each output tested.
11 Repeat steps (1) through (10) for the remaining outputs.
Output Readback Accuracy
+6V IO ± 10 mA
+25V IO ± 4 mA
-25V IO ± 4 mA
Output Programming Accuracy
+6V 5 A ± 20 mA
+25V 1 A ± 5.5 mA
-25V 1 A ± 5. 5 mA
Output Readback Accuracy
+6V IO ± 20 mA
+25V IO ± 5.5 mA
-25V IO ± 5.5 mA