Output Programming Range (maximum programmable values)

 

 

+6V Output

+25V Output

-25V Output

Voltage

0 to 6.18 V

0 to 25.75 V

0 to -25.75 V

Current

0 to 5.15 A

0 to 1.03 A

0 to 1.03 A

Temperature Coefficient, ±(% of output + offset)

Maximum change in output/readback per °C after a 30-minute warm-up

 

+6V Output

+25V Output

-25V Output

Voltage

0.01% + 2 mV

0.01%

+ 3 mV

0.01%

+ 3 mV

Current

0.02% + 3 mA

0.02%

+ 0.5 mA

0.02%

+ 0.5 mA

Stability, ±(% of output + offset)

Following a 30-minute warm-up, change in output over 8 hours under constant load, line, and ambient temperature

 

+6V Output

+25V Output

-25V Output

Voltage

0.03% + 1 mV

0.02% + 2 mV

0.02% + 2 mV

Current

0.1% + 3 mA

0.05% + 1 mA

0.05% + 1 mA

Voltage Programming Speed

Maximum time required for output voltage to settle within 1% of its total excursion (for resistive load). Excludes command processing time.

 

+6V Output

+25V Output

-25V Output

Full load Up

11 msec

50 msec

50 msec

Full load Down

13 msec

45 msec

45 msec

No load Up

10 msec

20 msec

20 msec

No load Down

200 msec

400 msec

400 msec

Isolation

The 0-6V supply is isolated from the ±25V supply up to ±240 Vdc. Maximum isolation voltage from any terminal to chassis ground ±240 Vdc.

AC Input Ratings (selectable via rear panel selector)

std

115 Vac ± 10%, 47 to 63 Hz

opt 0E3

230 Vac ± 10%, 47 to 63 Hz

opt 0E9

100 Vac ± 10%, 47 to 63 Hz

Maximum Input Power 350 VA with full load