Chapter 2 Quick Start
To store and recall the instrument state
34
To store and recall the instrument state
You can store up to three different operating states in non-volatile memory.
This also enables you to recall the entire instrument state with just a few key
presses from the front panel.
The memory locations are supplied from the factory for front panel operation
with the following states: display and knob selection for +6V output; *RST
values of voltage and current limits for three outputs; output disabled; and
track off state. *RST va lues for +6V supply are 0 V and 5 A and 0 V and 1 A
for the ±25V supplies.
The following steps show you how to store and recall an operating state.
1 Set up the power supply for the desired operating state.
The storage feature "remembers" the display and knob selection state, the limit
values of voltage and current for three outputs, output on/off state, and track
mode state.
2 Turn on the storage mode.
Three memory locations (numbered 1, 2, and 3) are available to store the
operating states. The operating states are stored in non-volatile memory and
are remembered when being recalled.
This message appears on the display for approximately 3 seconds.
3 Store the operating state in memory location "3".
Turn the knob to the right to specify the memory location 3.
To cancel the store operation, let the display time-out after 3 seconds or press
any other function key except the key. The power supply returns to
the normal operating mode and to the function pressed.

Store 1

Store 3

Store
Store