Store
Chapter 2 Quick Start
To store and recall the instrument state
To store and recall the instrument state
You can store up to three different operating states in
The memory locations are supplied from the factory for front panel operation with the following states: display and knob selection for +6V output; *RST values of voltage and current limits for three outputs; output disabled; and track off state. *RST values for +6V supply are 0 V and 5 A and 0 V and 1 A for the ±25V supplies.
The following steps show you how to store and recall an operating state.
1Set up the power supply for the desired operating state.
The storage feature "remembers" the display and knob selection state, the limit values of voltage and current for three outputs, output on/off state, and track mode state.
2Turn on the storage mode.
Three memory locations (numbered 1, 2, and 3) are available to store the operating states. The operating states are stored in
Store 1
This message appears on the display for approximately 3 seconds.
3Store the operating state in memory location "3". 
Turn the knob to the right to specify the memory location 3.
Store 3
To cancel the store operation, let the display
34