Chapter 3 Calibration Procedures

Constant Current (CC) Verifications

CC Line Regulation

This test measures the change in output current that results from a change in ac line voltage from the minimum value (10% below the nominal input voltage) to the maximum value (10% above nominal voltage).

1Turn off the power supply and connect the output to be tested as shown in Figure 3-1with the digital voltmeter connected across the current monitoring resistor (RM).

2Connect the ac power line through a variable voltage transformer.

3Turn on the power supply and select the desired output to be tested using the meter selection key on the front panel. Enable the outputs and set the display to the limit mode. When the display is in the limit mode, program the voltage to the maximum programmable value and the current to the full-scale value (see Table 3-2).

4Operate the electronic load in constant voltage mode and set its voltage to the full scale value of the output to be tested (see Table 3-2). Check that the CC annunciator remains lit. If not lit, adjust the load so that the output voltage drops slightly until the CC annunciator lights.

5Adjust the transformer to low line voltage limit (104 Vac for nominal 115 Vac, 90 Vac for nominal 100 Vac, or 207 Vac for nominal 230 Vac). Record the output current reading by dividing the voltage reading on the digital voltmeter by the resistance of the current monitoring resistor.

6Adjust the transformer to 10% above the nominal line voltage (110 Vac for a 100 Vac nominal input, 127 Vac for a 115 Vac nominal input or 253 Vac for a 230 Vac nominal input). Record the current reading again by dividing the voltage reading on the digital voltmeter by the resistance of the current monitoring resistor. The difference between the current readings in step (5) and (6) is the load regulation current. The difference of the readings should be within the limit specified below for each output tested.

Output

Difference

 

 

+6V

0.75 mA

 

 

+25V

0.35 mA

 

 

-25V

0.35 mA

 

 

7Repeat steps (1) through (6) for the remaining outputs.

57

3

Page 58
Image 58
Agilent Technologies E3631A manual CC Line Regulation