Chapter 3 Calibration Procedures

General Calibration/Adjustment Procedure

General Calibration/Adjustment Procedure

The calibration procedures from the front panel are described in this section. For output voltage calibration, disconnect all loads from the power supply and connect a DVM across the output terminals to be calibrated. For output current calibration, disconnect all loads from the power supply, connect an appropriate current monitoring resistor across the output terminals to be calibrated, and connect a DVM across the terminals of the monitoring resistor.

The following table shows calibration parameters and points which should be used to calibrate the output voltage and current for each of the three outputs.

Table 3-3. Parameters for Calibration

Output

Calibration Parameter

Voltage/

Calibration Point

Current

mnemonic

 

 

 

 

 

 

 

CAL SETUP 1

Voltage

V LO

 

 

+6V output

V HI

 

 

 

 

 

CAL SETUP 2

Current

I LO

 

 

 

 

I HI

 

 

 

 

 

 

 

 

CAL SETUP 3

Voltage

V LO

 

 

+25V output

V HI

 

 

 

 

 

CAL SETUP 4

Current

V LO

 

 

 

 

I HI

 

 

 

 

 

 

 

 

CAL SETUP 5

Voltage

V LO

 

 

-25V output

V HI

 

 

 

 

 

CAL SETUP 6

Current

I LO

 

 

 

 

I HI

 

 

 

 

 

 

 

You can terminate any CAL SETUP without changing its calibration constants by turning off power.

66

Page 67
Image 67
Agilent Technologies E3631A manual General Calibration/Adjustment Procedure