Chapter 3 Calibration Procedures
General Calibration/Adjustment Procedure
66
General Calibration/Adjustment Procedure
The calibration procedures from the front panel are described in this section.
For output voltage calibration, disconnect all loads from the power supply and
connect a DVM across the output terminals to be calibrated. For output current
calibration, disconnect all loads from the power supply, connect an
appropriate current monitoring resistor across the output terminals to be
calibrated, and connect a DVM across the terminals of the monitoring resistor.
The following table shows calibration parameters and points which should be
used to calibrate the output voltage and current for each of the three outputs.
Table 3-3. Parameters for Calibration
Note You can terminate any CAL SETUP without changing its calibration constants by
turning off power.
Output Calibration Parameter Voltage/
Current
Calibration Point
mnemonic
+6V output
CAL SETUP 1 Voltage V LO
V HI
CAL SETUP 2 Current I LO
I HI
+25V output
CAL SETUP 3 Voltage V LO
V HI
CAL SETUP 4 Current V LO
I HI
-25V output
CAL SETUP 5 Voltage V LO
V HI
CAL SETUP 6 Current I LO
I HI