Chapter 3 Calibration Procedures
Performance Test Record for Agilent E3631A
CC Performance Test Record
Test Description | Actual | Specifications | ||
|
|
| ||
Result | Upper Limit |
| Lower Limit | |
|
| |||
|
|
| ||
|
|
|
|
|
CC Programming Accuracy @ 0 amps (IO) |
|
|
|
|
+6V supply |
| +0.0100 A |
| |
+25V supply |
| +0.0040 A |
| |
| +0.0040 A |
| ||
|
|
|
|
|
CC Readback Accuracy @ 0 amps |
|
|
|
|
+6V supply |
| IO + 0.0100 A |
| IO - 0.0100 A |
+25V supply |
| IO + 0.0040 A |
| IO - 0.0040 A |
| IO + 0.0040 A |
| IO - 0.0040 A | |
CC Programming Accuracy @ Full Scale (IO) |
|
|
|
|
+6V supply |
| 5.0200 A |
| 4.9800 A |
+25V supply |
| 1.0055 A |
| 0.9945 A |
| 1.0055 A |
| 0.9945 A | |
|
|
|
|
|
CC Readback Accuracy @ Full Scale |
|
|
|
|
+6V supply |
| IO + 0.0200 A |
| IO - 0.0200 A |
+25V supply |
| IO + 0.0055 A |
| IO - 0.0055 A |
| IO + 0.0055 A |
| IO - 0.0055 A | |
CC Load Regulation |
| Maximum change: |
| |
+6V supply |
| <0.75 mA |
| |
+25V supply |
| <0.35 mA |
| |
| <0.35 mA |
| ||
|
|
|
| |
CC Line Regulation |
| Maximum change: |
| |
+6V supply |
| <0.75 mA |
| |
+25V supply |
| <0.35 mA |
| |
| <0.35 mA |
| ||
|
|
|
|
|
CC Ripple/Noise |
|
|
|
|
+6V supply |
| <2 mA |
| |
+25V supply |
| <0.5 mA |
| |
| <0.5 mA |
| ||
|
|
|
|
|
60