Chapter 3 Calibration Procedures

General Calibration/Adjustment Procedure

Calibrate

Calibrate

Connect an appropriate shunt (see Table 3-1 on page 42) across the +6V supply's output terminals, and connect a digital voltmeter across the shunt resistor for the current calibration.

10Select the first current calibration point for the +6V supply.

i lo +0.100 a

The display shows the first current calibration point for the +6V supply.

11Read the DVM and change the first current value on the display to match the computed current (DVM reading ¸ by shunt resistance).

For example, if the computed value is 0.099 A, adjust the current to 0.099 A using the knob and arrow keys.

Notice that you should wait for the DVM reading to stabilize for accurate calibration.

i lo +0.099 a

12Pressing the "Calibrate" key saves the change and selects the second current calibration point for the +6V supply.

i hi +4.750 a

If the entered number is within an acceptable range, an "ENTERED" message appears for one second. The display now shows the second current calibration point for the +6V supply. If the entered number is not correct, an "INVALID DATA" message appears for one second and the display shows the first current calibration point again.

69

3

Page 70
Image 70
Agilent Technologies E3631A manual Lo +0.100 a, Lo +0.099 a, Hi +4.750 a