Chapter 2 Quick Start

To use the power supply in constant voltage mode

Power

Output On/Off

Display Limit

To use the power supply in constant voltage mode

To set up the power supply for constant voltage (CV) operation, proceed as follows.

For each step, use the keys shown on the left margin.

1Connect a load to the desired output terminals.

With power-off, connect a load to the desired output terminals.

2Turn on the power supply.

The power supply will go into the power-on / reset state; all outputs are disabled (the OFF annunciator turns on); the display is selected for the +6V supply (the +6V annunciator turns on); and the knob is selected for voltage control.

3Enable the outputs.

The OFF annunciator turns off and the +6V and CV annunciators are lit. The blinking digit can be adjusted by turning the knob. Notice that the display is in the meter mode. "Meter mode" means that the display shows the actual output voltage and current.

To set up the power supply for +25V supply or -25V supply operation, you should press the +25V or -25Vkey to select the display and adjust for +25V supply or -25V supply before proceeding to the next step.

4Set the display for the limit mode.

Notice that the Lmt annunciator blinks, indicating that the display is in the limit mode. When the display is in the limit mode, you can see the voltage and current limit values of the selected supply.

In constant voltage mode, the voltage values between meter mode and limit mode are the same, but the current values are not. Further if the display is in the meter mode, you cannot see the change of current limit value when adjusting the knob. We recommend that you should set the display to "limit" mode to see the change of current limit value in constant voltage mode whenever adjusting the knob.

2

29

Page 30
Image 30
Agilent Technologies E3631A manual To use the power supply in constant voltage mode, Set the display for the limit mode