Chapter 3 Calibration Procedures

Constant Voltage (CV) Verifications

5Readback the output voltage of the selected output over the remote interface by sending the command:

MEAS:VOLT? {P6VP25VN25V}

6Record the value displayed on the controller. This value should be within the limits specified below for each output tested.

Output

Programming Accuracy

 

 

+6V

DVM ± 5 mV

 

 

+25V

DVM ± 10 mV

 

 

-25V

-(DVM ± 10 mV)

 

 

7Program the selected output's voltage to full scale value by sending the commands.

VOLT 6.0 for the +6V supply or

VOLT 25.0 for the ± 25V supply

8Record the output voltage reading on the digital voltmeter (DVM). The readings should be within the limits specified below for each output tested.

Output

Programming Accuracy

 

 

+6V

6 V ± 11 mV

 

 

+25V

25 V ± 32.5 mV

 

 

-25V

-(25 V ± 32.5 mV)

 

 

9Readback the output voltage of the selected output over the remote interface by sending the command:

MEAS:VOLT? {P6VP25VN25V}

10Record the value displayed on the controller. This value should be within the limits specified below for the output to be tested.

Output

Programming Accuracy

 

 

+6V

DVM ± 11 mV

 

 

+25V

DVM ± 22.5 mV

 

 

-25V

-(DVM ± 22.5 mV)

 

 

11Repeat steps (1) through (10) for the remaining outputs.