Fisher & Paykel 517735 service manual Phase 4 Motor Controller Fault, Primary Source- Pump

Models: 517735

1 92
Download 92 pages 53.84 Kb
Page 60
Image 60

61.(00111101) Phase 4 - Motor Controller Fault

The Motor Controller has been attempting to perform an internal EEPROM write. Prior to writing the Motor Controller has tested the 15 Volt supply and found that it is below the safety level for writing EEPROM and has reported this to the display. This may be due to mains transients at the time of write or due to a faulty Motor Controller.

Primary Source - Motor Controller Action - Replace Motor Controller

62.(00111110) Phase 5 - Pump Fault

The Motor Controller has detected an excessive pump current.

Primary Source- Pump

Action - Replace Pump. NB. Just checking the resistance of the pump (approx. 33 ohms cold to 40 ohms hot) is NOT reliable as the pump inductance is also important.

Secondary Source - Motor Controller Action - Replace Motor Controller.

63.(00111111) Phase 5 – Motor Controller Fault

The Motor Controller has detected an internal communications problem between its main control system and the pump control system.

Primary Source - Motor Controller Action - Replace Motor Controller.

64.(01000000) Phase 5 – Pressure Transducer Out of Range Error Refer to Fault code 66 for action.

65.(01000001) Phase 5 – Pressure Transducer Out of Range Error Refer to Fault code 66 for action.

66.(01000010) Phase 5 – Pressure Transducer Out of Range Error

The Motor Controller has received signals from the Pressure Transducer outside normal values.

Primary Source - Motor Controller Action - Replace Motor Controller

81 – 89. (010xxxx) Phase 2 to 3 - Display / Motor Controller Comms. Error

Refer to Fault code 106 for action.

- 55 -

Page 60
Image 60
Fisher & Paykel 517735 service manual Phase 4 Motor Controller Fault, Primary Source- Pump