rametersavailable for the current conversion type, an empty string is returned.
The list of parameters depends on the selected conversiontype. Characteriza-
tion parameters for the variousconversion types are listed in Table 8 on page
124, ConversionTypes and Parameters.
8.4.4.6 CALCulate<chn>:CONVert:PARameter:VALue? [<param>|ALL]
Returns the valueof the specified conversion parameter for the specified
channel.
CALC2:CONV:PAR:VAL? RTPW
100.0145
The CALCulatesuffix, <chn>, specifies the channel (1 to 4) or memory chan-
nel (5 to 14). The <param> parameter is the name of the parameter as givenin
Table8, Conversion Types and Parameters. If the <param> parameter does not
apply to theselected conversion type, an error -221, “Settings Conflict”, is
generated. Specifying the ALL parameter returns the name and valueof all pa-
rameters associated with the specified channel in a comma delimited list. If no
parameter is specified,the ALL parameter is assumed. The format of there-
sponse consistsof a parameter name, enclosed in double quotes, followed by its
value(i.e. "param1",value1,"param2",value2,"param3",value3,...).
8.4.4.7 CALCulate<chn>:CONVert:PARameter:VALue<param>,<num>[,<
param>,<num>...]
Sets thevalue of one or more conversion parameters for the specified channel.
CALC2:CONV:PAR:VALRTPW, 100.0145, A4, 0.0045
The CALCulatesuffix, <chn>, specifies the channel (1 to 4) or memory chan-
nel (5 to 14). The <param> parameter is the name of the parameter as givenin
Table8, Conversion Types and Parameters. If the <param> parameter is not ap-
plicable to the current conversiontype, an error -221, “Settings Conflict”, is
generated. The <num> parameter is the desired value of the parameter. Multiple
parameterscan be set using a single command by separating the <param> and
<num> prametersby commas.The order ofthe parameters is not important but
the valuemust immediately follow the parameter name.
This command may be password-protected. Toaccess this command:
If the SYST:PASS:CONV? command returns a 0 (probe password-protec-
tion is disabled), the password is not required.
If the SYST:PASS:CONV? command returns a 1 (probe password-protec-
tion is enabled), the password must be successfully entered using the
SYST:PASS:CEN<pass> command (see Section 9.4.11, Password Com-
mands).
8.4.4.8 CALCulate:CONVert:PRINt [<chn>|ALL [,<port>]]
Prints the conversionparameters for the specified channel.
95
8 Digital Communications Interface
Commands