bottomportion of the display shows the current display settings. This function
is thesame as the CHANNEL MODE menu DISPLAY OPTIONS function.
The RESOL parameter determines the number of decimal places (0 through 4
or AUTO)to be displayed for the temperature measurements. If AUTO is se-
lected, the instrument automatically determines the number of decimal places
to displaydepending on the conversion type.
The DECIMAL parameter is for selecting the character used as the decimal
separator.The choices are period (.) or comma (,).
The LAMP parameter specifiesthe brightness of the display (OFF,LOW,
MED, HIGH).Enabling this parameter causes the battery to discharge faster.
Press the LR buttonsto select the parameter setting. Use the UD buttons
to scroll between thep arameters. Pressthe Enter buttontosavethenewset
-
ting.Press the Ex it buttontocancelandtoexittothemenu.
7.4 Logging Menu
The LOGGING menu providesfunctions for logging measurements on demand
and automatically,displaying log statistics, editing data labels, and r esetting the
data labelsto their default values. The submenus and functions that appear in
51
7 Menu Functions
Logging Menu
1
24.159C
3
24.142C
2
24.634C
4
24.015C
RESOL: 3
DECIMAL:
.
LAMP: OFF
EXIT
ENTER / MENU UNIT CONTRAST
1529
CHUB E-4THERMOMETER READOUT
Figure24 Display Options