Press ENTER to continue…
Press the Enter buttonto confirm and to continue. The display is remains un-
changed and allenabled channels remain enabled channels. Press the Exit but-
ton to cancel.
When the measure period isreset to 1 second or greater, the following message
is displayedindicating that the 1529 is returning to normal operation.
FULL ACCURACY.
The DISPLAYFUNCTIONS function of the CHANNEL menu reappears and
the CHANNELMODE function is available.
Note: Accuracyin the Fast Measurement Mode is affected by ambient condi-
tions, changesin measured temperature, higher measured temperature, and the
timeduration from using the AUTO-CAL function. To reduce the uncertainty
in the FastMeasurement Mode, use the 1529 in a controlled ambient environ-
ment, in a verystable bath or d ry block at lowertemperatures, and u seth e
AUTO-CALfunction often.
7.1.2 Enable ChannelThe ENABLECHANNEL function enables and disables measuring of each
channel. If a channel is off,its measurement is not displayed on the top portion
of the display in either scan or simultaneous display modes. When this function
is selected, the bottom portion of the display showsthe on/off status of each
configured channel and allows the setting to be changed. Each configured chan-
nel is identifiedby its channel number.
Use theUD buttons to select the channel to change. The LR buttons are
used to change the setting.Press the Enter button to save the setting. Press and
1529 Chub-E4 Thermometer Readout
User’s Guide
30
1
24.159C
3
24.142C
2
24.634C
4
24.015C
CHAN 1: ON
CHAN 2: ON
CHAN 3:
ON
CHAN 4: ON
EXIT
ENTER / MENU UNIT CONTRAST
1529
CHUB E-4THERMOMETER READOUT
Figure9 Enable Channel