The systemreset does the following:
Sets the measure period to 1 second.
Sets channel 1 ON, all others OFF.
Sets the channel mode to SCAN.
Sets moving average to 1.
Sets printing to all ports OFF.
Stops auto logging.
Clears the statistical functions.
Sets units to C.
Sets the resolution to AUTO.
Resetdoes not affectany of the probe characterization parameters, logged data
or channel calibration parameters.
1529 Chub-E4 Thermometer Readout
User’s Guide
76