entered, the bottom portion of the display allows the user to select the channel
to view.
Each channel of the1529 is calibrated independently and therefore each chan-
nel has its own set of calibration parameters which are displayed after the CAL
CH parameter isentered.
Use theLR buttons to select the channel. Press the Enter button to view the
calibration parameters for the selected channel. Press the Exit button to cancel
and to exitto the menu.
ForPRT/thermistor channels, the parameters C0, C100, C400, C0K, C10K,
C100K, DATE,and DUE are displayed.
The C0, C100, and C400 parameters apply to the PRTprobe types and
100Ωresistance range. Use these parameters to adjust the measurement at
0, 100 and 400Ωrespectively. These values are in ohms.
The C0K, C10K, and C100K parameters apply to the thermistor probe
types and 10KΩresistance range. Use these parameters to adjust the mea-
surement at 0K,10K, and 100KΩrespectively. These values are inKohms.
The DATE parameter is the date the current calibration was performed.
The DUE parameter is the date the current calibration expires.
Forthermocouple channels, the parameters C0, C100, CRJ, DATE, and DUE
are displayed.
The C0 and C100 parameters adjust the measurement at 0 and 100 mV re-
spectively. These values are in mV.
The CRJ parameter adjusts the reference junction temperature measure-
ment. This parameter is in °C.